Fisher 244EH, 244ER manual Configuration Software, Multichannel Installations, System Requirements

Models: 244EH 244ER

1 64
Download 64 pages 54.53 Kb
Page 19
Image 19
Multichannel Installations

Multichannel Installations

Several transmitters can be connected to a single master power supply as shown in Figure 2-9.In this case, the system may be grounded only at the negative power supply terminal. In multichannel installations where several transmitters depend on one power supply and the loss of all transmitters would cause operational problems, consider a back-up battery or power supply that cannot be interrupted. The diodes shown in Figure 2-9prevent unwanted charging or discharging of the back-up battery.

Figure 2-9. Multichannel Installations

 

RLead

Transmitter

RLead

No. 1

 

Transmitter

RLead

No. 2

Backup Battery

Readout or

dc

Controller

Power

No.1

Supply

Readout or

 

Controller

 

No. 2

To Additional Transmitters

3044-0131A

Configuration Software

System Requirements

 

For the configuration software to function properly, the following

 

minimum PC requirements must be observed.

 

• 80386 or faster processor

 

• 1 MB free disk space

 

3.5-inch floppy disk drive

 

• Mouse or other pointing device

 

• 256 color display (recommended settings are 16-bit color at

 

1024 × 728 screen resolution)

 

• Microsoft Windows 3.1, Windows for Workgroups 3.11,

 

Windows 95, or Windows NT.

 

The configuration software is available in English, French, German,

 

Italian, Spanish, Chinese, Japanese, and Korean language versions.

 

The configuration software automatically selects the appropriate

 

language based on the language version of Windows.

NOTE

Close all currently running applications before starting the installation procedure.

NOTE

Windows 98 is not compatible with this software.

2-9

Page 19
Image 19
Fisher 244EH, 244ER manual Configuration Software, Multichannel Installations, System Requirements