Accuracy

Accuracy

TABLE A-2. Model 244E Input Options and Accuracy.

Sensor

Sensor

Input

Recommended

 

Options

Reference

Ranges

Min. Span(1)

Accuracy (whichever is greater)

2-, 3-, 4-wire RTDs

 

°C

°F

°C

°F

 

 

 

 

 

 

 

 

Pt 100

IEC 751, 1995 (α = 0.00385)

–200 to 850

–328 to 1562

10

18

0.05% of span + 0.15 °C or 0.2 °C

Pt 100

JIS 1604, 1981 (α = 0.003916)

–200 to 645

–328 to 1093

10

18

0.05% of span + 0.15 °C or 0.2 °C

 

 

 

 

 

 

 

Pt 200

IEC 751, 1995 (α = 0.00385)

–200 to 850

–328 to 1562

10

18

0.01% of span or 0.4 °C

PT 500

IEC 751, 1995 (α = 0.00385)

–200 to 850

–328 to 1562

10

18

0.01% of span or 0.3 °C

 

 

 

 

 

 

 

Pt 1000

IEC 751, 1995 (α = 0.00385)

–200 to 300

–328 to 572

10

18

0.01% of span or 0.3 °C

Ni 120

Edison Curve No. 7

–70 to 300

–94 to 572

10

18

0.01% of span or 0.2 °C

 

 

 

 

 

 

 

Cu 10

Edison Copper Winding No. 1

–50 to 250

–58 to 482

10

18

0.5% of span or 1.5 °C

Thermocouples(2)

 

 

 

 

 

 

Type B(3)

NIST Monograph 175, IEC 584

100 to 1820

212 to 3308

25

40

0.2% of span or 1.0 °C

Type E

NIST Monograph 175, IEC 584

–50 to 1000

–58 to 1832

25

40

0.1% of span or 0.5 °C

 

 

 

 

 

 

 

Type jJ

NIST Monograph 175, IEC 584

–180 to 760

–292 to 1400

25

40

0.1% of span or 0.5 °C

Type K

NIST Monograph 175, IEC 584

–180 to 1372

–292 to 2502

25

40

0.1% of span or 1.0 °C

 

 

 

 

 

 

 

Type N

NIST Monograph 175, IEC 584

–200 to 1300

–328 to 2372

25

40

0.1% of span or 1.0 °C

Type R

NIST Monograph 175, IEC 584

0 to 1768

32 to 3214

25

40

0.1% of span or 1.0 °C

 

 

 

 

 

 

 

Type S

NIST Monograph 175, IEC 584

0 to 1768

32 to 3214

25

40

0.1% of span or 1.0°C

Type T

NIST Monograph 175, IEC 584

–200 to 400

–328 to 752

25

40

0.1% of span or 0.5 °C

 

 

 

 

 

 

 

DIN Type L

DIN 43710

–200 to 900

–328 to 1652

25

40

0.1% of span or 0.5 °C

DIN Type U

DIN 43710

–200 to 600

–328 to 1112

25

40

0.1% of span or 0.5 °C

 

 

 

 

 

 

 

Type

ASTME 988-96

0 to 2000

32 to 3632

25

40

0.1% of span or 1.0 °C

W5Re/W26Re

 

 

 

 

 

 

Millivolt Input

 

–10 to 100 mV

 

3 mV

0.025 mV + 0.003% of span

 

 

 

 

2-, 3-, 4-wire Ohm Input

0 to 2000 ohms

20 ohm

0.75 Ω + 0.03% of span

 

 

 

 

 

 

 

(1)No minimum or maximum span restrictions within the input ranges. Recommended minimum span will hold noise within accuracy specification with damping at zero seconds.

(2)Total digital accuracy for thermocouple measurement: sum of digital accuracy +0.5 °C (cold junction accuracy)

(3)Accuracy for NIST Type B thermocouple is ±3.0 °C from 100 to 300 °C.

When using a Pt 100 (α = 0.00385) sensor input with a 75 to 150 °C range, the accuracy will be 0.05% of span + 0.15 °C or 0.20 °C, whichever is greater. Sample Calculation:[0.0005 (150-75) + 0.15] = 0.19 °C, which is less than 0.20 °C, so the accuracy equals 0.20 °C.