3 Specifications
3.1Mechanical
Dimensions: | 140 mm x 173 x 135 (without chamber fitted) |
Weight: | 2.8 kg (without chamber fitted) |
| Approx. 3.1 kg (with chamber fitted, and filled with water) |
3.2 | Electrical |
|
|
| |
|
|
|
|
| |
MR850 Model Number |
| Supply Voltage |
| Supply Current | |
MR850Axx |
| 230 V~ |
| 1.0 A Max | |
MR850Pxx |
| 127 V~ |
| 1.8 A Max | |
MR850Jxx |
| 115 V~ |
| 2.0 A Max | |
MR850Gxx |
| 100 V~ |
| 2.4 A Max | |
Supply Frequency: | 50 or 60 Hz, Sine Wave |
| |||
Heater Plate Capacity | 150 W at nominal mains voltage | ||||
HP Thermal Cutout: | 118 ± 6 °C |
| |||
Heater Wire Supply: | 22 ± 5 V~, 2.73 A Max, 50 or 60 Hz | ||||
Maximum Heater Wire Load: | 8.0 Ω. |
|
3.3Temperature Range
3.3.1Heater Wire Mode
Invasive Mode: | Chamber Set Point: 35.5 to 37 °C |
| Chamber Set Point: 35.5 to 40 °C (versions 5.33, 5.34, 5.45, 5.70, |
| 6.00, 7.00, 7.21). |
| Chamber Set Point: 35.5 to 42 °C (versions 7.14, 7.17 & 7.22). |
| Airway Set Point: 35 to 40 °C |
Chamber Set Point: 31 °C | |
| Chamber Set Point: 31 to 34 °C (versions 5.33, 5.34, 5.45, 5.70, |
| 6.00). |
| Chamber Set Point: 31 to 36 °C (versions 7.00, 7.14, 7.17, 7.21, |
| 7.22). |
| Airway Set Point: 28 to 34 °C |
3.3.2Non Heater Wire Mode
Invasive Mode: | Airway Set point: 37 | °C (chamber temperature limited to 66 °C) |
Airway Set point: 31 | °C (chamber temperature limited to 66 °C) | |
Display: | Three digit, 14 mm, 7 segment LED | |
| Range: 10 to 70 °C |
|
Accuracy: | ± 0.3 °C (in 25 to 45 | °C temperature range) |
3.3.3Alarm Parameters
High Temperature Alarm: | Causes an immediate, audible and visible alarm at a displayed | |
|
| temperature of 41 °C or if the airway temperature exceeds 43 °C |
|
| (see section 4.3) |
Temperature Alarm: | Invasive Mode: | |
|
| After 10 minutes @ 29.5 °C causes an audible and visible alarm. |
|
| After 60 minutes @ 34.5 °C causes an audible and visible alarm |
|
| (see section 4.5) |
|
| NOTE: The temperature indicator lights if the displayed |
|
| temperature drops below 35.4 °C, initially providing a temperature |
|
| warning. |
10 | MR850 Technical Manual | * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) |