Symptom | Corrective Action | Reference | |
A low temperature | 1. | Make sure the humidifier has had time to warm up and that |
|
is shown on the |
| there is sufficient gas flow through the breathing circuit. |
|
humidifier’s |
|
|
|
2. | The humidifier cannot maintain temperature. If the | Section 4.5 | |
display, with no |
| temperature indicator is also on then an alarm will occur |
|
audible alarm |
| eventually. |
|
| 3. | Humidifier or probe faulty. Complete humidifier and probe | Section 8 |
|
| performance test. Replace probe or service humidifier as |
|
|
| required. |
|
High displayed | 1. | Flow has recently been changed, allow 30 minutes for |
|
temperature, no |
| temperature to stabilise |
|
temperature alarm | 2. | Manual or automatic humidity compensation is active | Section 4.1.5 |
Temperature |
| The gas flow rate may have suddenly changed. Monitor | Section 8 |
indicator flashes, |
| the displayed temperature, if the temperature does not fall |
|
with audible alarm, |
| rapidly then remove humidifier from patient, and complete a |
|
and a high |
| performance test on the humidifier, and temperature / flow |
|
temperature is |
| probe. Replace probe or service humidifier as required. |
|
shown. |
|
|
|
6.2Technical Problems
Problems that cause the humidifier to malfunction without an audible alarm are discussed in this section.
Symptom | Corrective Action | Reference | |
See manual | 1. | Check that the mains supply is within specification. |
|
indicator lit (not |
| Provide adequate mains supply if required. |
|
flashing) | 2. | If ‘PtS’ is displayed. | Section 6.1 |
| 3. | Replace PCBs. | Section 7.2.3 |
Humidifier will not | 1. | Humidifier plugged into mains supply? |
|
turn on with on/off | 2. | Check that the mains supply is within specification. |
|
button, and no |
| Provide adequate mains supply if required. |
|
indicators are lit | 3. | Remove mains power and check the fuses. | Section 7.2.2 |
| |||
| 4. | Check continuity of mains power cord. Replace as | Section 7.2.7 |
|
| necessary. |
|
| 5. | Check the transformer windings are not open circuit. | Section 7.2.4 |
|
| Primary Winding: Red wires |
|
|
| Secondary 10v winding: White wires |
|
|
| Secondary 22v winding: Yellow wires |
|
|
| If any windings are open circuit, replace the transformer. |
|
| 6. | Replace PCBs. | Section 7.2.3 |
|
|
|
|
Unit fails |
| Send unit to a Fisher & Paykel Healthcare service | Section 7.2.3 |
calibration check |
| representative for calibration, or replace PCBs. |
|
Probe fails Probe |
| Replace probe. |
|
Accuracy Check |
|
|
|
Unit fails to reach | 1. | If the chamber outlet temperature was low, the heater plate | Section 7.2.5 |
temperature in |
| element should be checked. Replace element as |
|
humidifier control |
| necessary. |
|
check. (Section | 2. | If the airway temperature was low, replace breathing circuit |
|
8.1.4) |
| and |
|
| 3. | Replace PCBs. | Section 7.2.3 |
|
|
|
|