Symptom

Corrective Action

Reference

A low temperature

1.

Make sure the humidifier has had time to warm up and that

 

is shown on the

 

there is sufficient gas flow through the breathing circuit.

 

humidifier’s

 

 

 

2.

The humidifier cannot maintain temperature. If the

Section 4.5

display, with no

 

temperature indicator is also on then an alarm will occur

 

audible alarm

 

eventually.

 

 

3.

Humidifier or probe faulty. Complete humidifier and probe

Section 8

 

 

performance test. Replace probe or service humidifier as

 

 

 

required.

 

High displayed

1.

Flow has recently been changed, allow 30 minutes for

 

temperature, no

 

temperature to stabilise

 

temperature alarm

2.

Manual or automatic humidity compensation is active

Section 4.1.5

Temperature

 

The gas flow rate may have suddenly changed. Monitor

Section 8

indicator flashes,

 

the displayed temperature, if the temperature does not fall

 

with audible alarm,

 

rapidly then remove humidifier from patient, and complete a

 

and a high

 

performance test on the humidifier, and temperature / flow

 

temperature is

 

probe. Replace probe or service humidifier as required.

 

shown.

 

 

 

6.2Technical Problems

Problems that cause the humidifier to malfunction without an audible alarm are discussed in this section.

Symptom

Corrective Action

Reference

See manual

1.

Check that the mains supply is within specification.

 

indicator lit (not

 

Provide adequate mains supply if required.

 

flashing)

2.

If ‘PtS’ is displayed.

Section 6.1

 

3.

Replace PCBs.

Section 7.2.3

Humidifier will not

1.

Humidifier plugged into mains supply?

 

turn on with on/off

2.

Check that the mains supply is within specification.

 

button, and no

 

Provide adequate mains supply if required.

 

indicators are lit

3.

Remove mains power and check the fuses.

Section 7.2.2

 

 

4.

Check continuity of mains power cord. Replace as

Section 7.2.7

 

 

necessary.

 

 

5.

Check the transformer windings are not open circuit.

Section 7.2.4

 

 

Primary Winding: Red wires

 

 

 

Secondary 10v winding: White wires

 

 

 

Secondary 22v winding: Yellow wires

 

 

 

If any windings are open circuit, replace the transformer.

 

 

6.

Replace PCBs.

Section 7.2.3

 

 

 

 

Unit fails

 

Send unit to a Fisher & Paykel Healthcare service

Section 7.2.3

calibration check

 

representative for calibration, or replace PCBs.

 

Probe fails Probe

 

Replace probe.

 

Accuracy Check

 

 

 

Unit fails to reach

1.

If the chamber outlet temperature was low, the heater plate

Section 7.2.5

temperature in

 

element should be checked. Replace element as

 

humidifier control

 

necessary.

 

check. (Section

2.

If the airway temperature was low, replace breathing circuit

 

8.1.4)

 

and re-test.

 

 

3.

Replace PCBs.

Section 7.2.3

 

 

 

 

26 MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)