6 Troubleshooting
6.1Operational Problems
This section deals with faults that cause the humidifier to alarm. This may be caused by incorrect setup, faulty accessories, or a faulty humidifier. Refer below for troubleshooting.
Symptom | Corrective Action | Reference | ||
See manual | Record the displayed error code. | Section 6.3 | ||
indicator flashing |
|
|
| |
with audible alarm |
|
|
| |
See manual | 1. | If “PtS” is displayed on the front panel, then the production | Section 7.2.6 | |
indicator light |
| test mode has been accessed. Check the production test |
| |
permanently on, |
| button is not depressed or mode set by serial command. |
| |
without an audible | 2. | Humidifier faulty. Replace PCBs. | Section 7.2.3 | |
alarm |
|
|
|
|
Water Out indicator | 1. | Check that there is sufficient water in the chamber. Refill |
| |
flashes, |
|
| or replace chamber as necessary. Check that the water |
|
accompanied by an |
| bag is not empty, and the delivery tube is not kinked or |
| |
audible alarm |
| occluded. |
| |
|
| 2. | Check that the water level in the MR290 chamber is not |
|
|
|
| above the marked line. Replace chamber if the water is |
|
|
|
| above this line. |
|
|
| 3. | Check that the gas flow rate is within specification of the |
|
|
|
| humidifier and accessories being used. Adjust as |
|
|
|
| necessary. |
|
|
| 4. | Has condensate formed on the chamber probe? Dry |
|
|
|
| Probe and |
|
|
| 5. | Temperature probe faulty. Complete a temperature and | Section 8.2 |
|
|
| flow accuracy test on the probe. Replace probe as |
|
|
|
| required. |
|
|
| 6. | Humidifier faulty. Complete a performance test. Service | Section 8 |
|
|
| humidifier as required. |
|
Chamber Probe | 1. | Check that the chamber probe is inserted into the |
| |
alarm flashes |
| breathing circuit correctly, and that the breathing circuit is |
| |
accompanied by an |
| set up correctly. |
| |
audible alarm | 2. | Check that there is sufficient water in the chamber. Refill |
| |
|
|
| as necessary. Check that the water bag and delivery tube |
|
|
|
| are not kinked or occluded. |
|
|
| 3. | Ensure correct chamber is being used (refer Operating |
|
|
|
| Manual) |
|
|
| 4. | Check that the gas flow rate is within specification of the |
|
|
|
| humidifier and accessories being used. Adjust as |
|
|
|
| necessary. |
|
|
| 5. | Has condensate formed on the chamber probe? Dry |
|
|
|
| probe and |
|
|
| 6. | Temperature probe or humidifier faulty. Complete | Section 8 |
|
|
| performance tests. Replace probe or service humidifier as |
|
|
|
| required. |
|
Heater wire | 1. | Check that the heater wire adaptor is correctly plugged into |
| |
connector alarm |
| the humidifier along with the breathing circuit. |
| |
flashes, |
|
| NOTE: The short lead must connect to the inspiratory limb. |
|
accompanied by an | 2. | Replace breathing circuit, and |
| |
audible alarm |
| |||
3. | Replace heater wire adaptor, and check for intermittent |
| ||
|
|
| ||
|
|
| connections. |
|
|
| 4. | Humidifier faulty. Replace PCBs. | Section 7.2.3 |
24 | MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) |