| Invasive Mode: Airway temperature < 29.5 °C causes an audible |
| and visible alarm. |
| |
| audible and visible alarm. |
Sound Pressure Level: | Alarms exceed 50 dBA @ 1 m. |
3.4Performance
Invasive Mode: | Flow up to 60 LPM, humidity output >33 mg/L |
Flow up to 120 LPM, humidity output >10 mg/L | |
| NOTE: Performance results with RT100 breathing circuit |
Maximum System | 20 kPa, gas leakage at max. pressure <100 mL/minute. |
Operating Pressure: |
|
Less than 30 minutes. | |
Recommended ambient | 18 to 26 °C |
temperature range: |
|
CAUTION: If operating outside the recommended ambient temperature range, consult your local Fisher & Paykel Healthcare representative or refer section 4.1.5.
3.5Transport and Storage Environmental Conditions
Transport Conditions: | |
Storage Conditions: |
3.6Standards and Approvals
AS/NZS 3200.1.0, CAN/CSA 22.2 No.601.1, UL
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 11 |