5.2Safety Check
The unit should be tested to the current medical electrical standards for
CAUTION: Permanent damage to this humidifier will result if the serial port is used as a ground point during electrical safety testing.
NOTE: The correct ground test point location is on the heater plate front underside edge, as shown in Figure 5.1., where the insulating anodizing layer has been removed.
Figure 5.1 Showing the correct location of ground test point on the heater plate