5.2Safety Check

The unit should be tested to the current medical electrical standards for in-house testing for each specific country (example, refer to AS/NZS 3551 for Australia and New Zealand).

CAUTION: Permanent damage to this humidifier will result if the serial port is used as a ground point during electrical safety testing.

NOTE: The correct ground test point location is on the heater plate front underside edge, as shown in Figure 5.1., where the insulating anodizing layer has been removed.

Figure 5.1 Showing the correct location of ground test point on the heater plate

22 MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)