4.4.2Temperature / Flow Probe Connector
This indicator will light if the temperature probe is not correctly plugged in, or the probe used is faulty. The humidifier tests for the following probe fault conditions:
•Temperature probe disconnected
•Chamber thermistor open or short circuit
•Airway thermistor open or short circuit
•Flow thermistor open or short circuit (shorted probe test)
•One thermistor shorted to another (shorted probe test)
•Flow calibration resistor open or short circuit (shorted probe test)
An alarm will be generated if any of the above faults are found, and the humidifier will remove power from all heating systems.
NOTE: the shorted probe tests and flow thermistor tests are only performed on
4.4.3Chamber Probe & Airway Probe
These indicators are used to show that either the chamber probe or airway probe is not inserted into the breathing circuit correctly. On
During periods of low or zero gas flow, the airway probe out alarm is disabled. As soon as flow is detected however, an airway probe test is initiated.
NOTE: For software versions 7.22 and 7.23 in
4.4.4Chamber or Airway Probe Alarm with Probe connector alarm
The humidifier checks to see if the temperature probe is faulty by testing for the following conditions:
•Chamber temperature has been greater than 50 °C for 20 minutes
•Chamber temperature is greater than 80 °C
•Airway temperature has been greater than 50 °C for 5 minutes
•Airway temperature is greater than 80 °C
If an apparent fault is found, the humidifier will give a temperature / flow probe connector alarm, and also indicate either the chamber or airway probe. The humidifier will stay in
18 MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)