4.1.2
Ventilator/Gas Supply
RT134 Breathing Circuit
Temperature Probe
Chamber
Probe
MR290 Chamber
Airway Probe
Water trap
MR850 Humidifier
Figure 4.2 Typical
In this application the MR850 maintains the airway temperature at the desired set point (invasive 37 °C or
Activating
Refer also to the diagnostic menu in section 6.4.
1Turn on the humidifier with NO heated circuit connected, and mute the heater wire alarm.
2Access the diagnostic menu by pressing the mute and mode buttons together for 1 second, the display should show two rows of dashes ‘= = =’. Releasing both buttons will allow the diagnostic menu to cycle automatically.
3Allow the menu to cycle through to “cct” – the circuit identification. Press mute to access this function.
4To enable
Deactivating
The simplest way to
4.1.3
If the humidifier detects a problem with its setup or operation it will alarm. Depending on the severity of the alarm condition, the humidifier will either remove all power from the heating systems, or enter
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 13 |