6.4.3Diagnostic Menu for Software Version 7.00 & 7.21
Display | Description |
| |
HC | Humidity Compensation (HC) algorithm |
| |
| Invasive mode, compensation range is 0.0 to 3.0 °C ( CSP = 37.0 to 40 °C ) |
| |
|
| ||
| By pressing the Mute and Mode buttons together for 1 second or pressing the Mute | ||
| and Power buttons together for 1 second the user can respectively move up or |
| |
| down through the settings listed below. The humidifier will confirm the change with | ||
| a |
| |
| ‘5.0’ = +5.0 °C of chamber compensation |
| |
| ‘4.0’ = +4.0 °C of chamber compensation |
| |
| ‘3.0’ = +3.0 °C of chamber compensation |
| |
| ‘2.0’ = +2.0 °C of chamber compensation |
| |
| ‘1.0’ = +1.0 °C of chamber compensation |
| |
| ‘0.0’ = +0.0 °C of chamber compensation (No compensation) |
| |
|
| ||
|
| Invasive mode: 0 to 3 °C in 0.5 °C steps |
|
|
|
| |
CSP | Chamber set point, in 0.1 °C resolution eg: 37.0 |
| |
| Invasive mode range 35.5 to 40 °C. |
| |
|
| ||
Cdc | Chamber Duty Cycle (%) |
| |
CHP | Chamber Power / Flow ratio (W/LPM) |
| |
|
|
| |
hP | Heater Plate Temperature (1 °C) |
| |
Flo | Gas Flow Rate (0.1 LPM) |
| |
|
| ||
FLr | Gas Flow Rate Range: |
| |
|
| ||
| “no” = No flow |
| |
| “Lo” = Low flow, ( < 3 LPM ) |
| |
| “In” | = Intermediate flow, ( 2 to 17 LPM ) |
|
| “Hi” | = High flow, ( > 13 LPM ) |
|
| “ | = Ventilated flow detected |
|
ASP | Airway temperature set point (0.1 °C) |
| |
|
|
| |
Adc | Airway Duty Cycle (%) |
| |
Cct | Connected Circuit heater identification: |
| |
| “S“ = Standard inspiratory heater connected |
| |
| “C” = Coaxial inspiratory heater connected |
| |
| “E” = Expiratory heater connected |
| |
|
| ||
LAS | Last Alarm State, the display will blank, and the humidifier’s last alarm will be shown | ||
| on the indicators. To clear LAS, press and hold the mute and mode buttons for 1 |
| |
| second. The humidifier will beep when LAS has been cleared. |
| |
LFS | Last Fault State, refer to section 6.3. To clear LFS, press and hold the mute and |
| |
| mode buttons for 1 second. The humidifier will beep when LFS has been cleared. |
| |
SoF | Software version |
| |
End | Press the mute button to cycle to the start of the menu, or the diagnostic menu will |
| |
| automatically exit after 6 seconds |
| |
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 31 |