7.2.6Installing New Software
NOTE: A software upgrade kit is required.
Some software may not be available in your country. Refer to your local Fisher & Paykel Healthcare representative for the appropriate part number:
| Single ROM pack | 32 ROM Pack |
GJU model | 043042459 | 043042461 |
JHU model | 043042458 | 043042460 |
All other models | 043042066 | 043041255 |
Figure 7.8 Showing location of U3
1.Open the case (section 7.2.1).
2.Remove ROM integrated circuit U3 (Figure 7.8) using appropriate tools.
3.Place new ROM into the empty socket, ensuring correct orientation.
4.Close the case (section 7.2.8)
5.Turn the humidifier upside down; use a non conducting tool to hold the production test button down. This button is accessible through the right air vent slot (when looking at the rear of the humidifier). See Figure 7.9.
WARNING:
Never use an electrically conducting tool to depress the production test button, as there are live mains voltages inside the humidifier.
Figure 7.9 Showing Location of the Production Test Button