6.4.5Diagnostic Menu for Software Version 7.22
Display | Description |
|
| |
HC | Humidity Compensation (HC) algorithm |
|
| |
| Note: HC is inactive while operating under |
| ||
| Invasive mode, compensation range is 0.0 to 5.0 °C ( CSP = 37.0 to 42 °C ) |
| ||
|
| |||
| By pressing the Mute and Mode buttons together for 1 second or pressing the Mute and Power buttons |
| ||
| together for 1 second the user can respectively move up or down through the settings listed below. The |
| ||
| humidifier will confirm the change with a |
| ||
| ‘5.0’ = +5.0 °C of chamber compensation |
|
| |
| ‘4.0’ = +4.0 °C of chamber compensation |
|
| |
| ‘3.0’ = +3.0 °C of chamber compensation |
|
| |
| ‘2.0’ = +2.0 °C of chamber compensation |
|
| |
| ‘1.0’ = +1.0 °C of chamber compensation |
|
| |
| ‘0.0’ = +0.0 °C of chamber compensation (NO compensation) |
| ||
| Invasive mode: 0 to 5 °C in 0.5 °C steps |
| ||
|
|
|
| |
|
|
|
| |
Cct | Connected breathing circuit identification: |
|
| |
| “S“ = Standard inspiratory heater connected |
|
| |
| “C” = Coaxial inspiratory heater connected |
|
| |
| “E” = Expiratory heater connected |
|
| |
|
| |||
|
|
| ||
| To enable non heater wire operation, press and hold both the mute and mode buttons simultaneously for 1 |
| ||
| second. The display will show ‘nhh’ and the humidifier will confirm with an audible beep. Ensure that no |
| ||
| heated breathing circuit is connected to the humidifier otherwise the humidifier won’t change operation. |
| ||
| To disable |
| ||
CSP | Chamber set point, in 0.1 °C resolution eg: 37.0 |
|
| |
| Invasive mode range 35.5 to 42 °C. |
|
| |
|
|
| ||
|
|
|
| |
Cdc | Chamber Duty Cycle (%) |
|
| |
| During selection of this menu and while a chamber probe out test is active the chamber probe indicator will |
| ||
| light. |
|
|
|
CHP | Chamber Power / Flow ratio (W/LPM) |
|
| |
hP | Heater Plate Temperature (1 °C) |
|
| |
|
|
|
| |
Flo | Gas Flow Rate (0.1 LPM) |
|
| |
|
| |||
|
|
|
| |
FLr | Gas Flow Rate Range: |
|
| |
|
| |||
| “no” = No flow, ( |
|
| |
| “Lo” = Low flow, ( < 3 LPM ) |
|
| |
| “In” | = Intermediate flow, ( 2 to 17 LPM ) |
|
|
| “Hi” | = High flow, ( > 13 LPM ) |
|
|
| “ | = Ventilated flow detected |
|
|
ASP | Airway temperature set point (0.1 °C) |
|
| |
|
|
|
| |
Adc | Airway Duty Cycle (%) |
|
| |
| During selection of this menu and while an airway probe out test is active the airway probe indicator will light. | |||
|
| |||
H2O | Water out number, used to detect the presence of chamber water, calculated from, chamber power / (heater | |||
| plate temp. – chamber temp.). |
|
| |
| During selection of this menu and while the water out number falls below a dry chamber threshold the water | |||
| out indicator will light. |
|
| |
LAS | Last Alarm State, the display will blank, and the humidifier’s last alarm will be shown on the indicators. To |
| ||
| clear LAS, press and hold the mute and mode buttons for 1 second. The humidifier will beep when LAS has | |||
| been cleared. |
|
| |
LFS | Last Fault State, refer to section 6.3. To clear LFS, press and hold the mute and mode buttons for 1 second. | |||
| The humidifier will beep when LFS has been cleared. |
| ||
SoF | Software version |
|
| |
|
|
| ||
End | Press the mute button to cycle to the start of the menu, or the diagnostic menu will automatically exit after 6 |
| ||
| seconds. |
|
| |
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 33 |