8.1.4Humidifier Voltage Calibration Check

This check is required when the humidifier's PCBs have been serviced or replaced, or if the mains transformer has been replaced.

Equipment Required:

An AC voltmeter, capable of measuring RMS mains voltage to ±0.5 % accuracy.

Suitable breathing circuit for MR850 (for example 900RT100).

MR850 Heater Wire Adaptor (for example 900MR800).

1.Connect a breathing circuit to the humidifier with the heater wire adaptor. Make sure the inspiratory limb is correctly plugged in, and that a temperature or calibration probe is not plugged in to the humidifier.

2.Safely connect the AC voltmeter to the mains supply.

NOTE: Connect the voltmeter in close proximity to the mains socket used to power the humidifier.

3.Enter the service menu (refer section 8.1.1).

4.Select service mode number 3 (Voltage Calibration Check) by pushing the mute button when '-3-' is displayed.

5.Calculate the percentage of the actual mains voltage (the AC voltmeter reading) to the nominal mains voltage for the humidifier model being tested.

NOTE: The nominal mains voltage is the voltage indicated on the left side of the humidifiers case, and will be either 230 V~, 127 V~, 115 V~, or 100 V~.

Percentage mains = 100 x [Actual mains voltage (RMS) / Nominal Mains voltage]

For example:

MR850 Model AEK :

Voltmeter Reading : Percentage mains: Round any decimal points:

Nominal Mains Voltage is 230 V~ 240.5 V~

100 x [240.5 / 230] = 104.5 %

105

6.Adjust the reading on the humidifier's display to read the same as the value calculated above. This is achieved by pushing the mode button to increase, or the power button to decrease the percent value on the humidifier's display.

7.Check that the mains voltage has remained constant. If necessary recalculate the percentage of actual mains voltage.

8.Store the new value in the humidifier's memory by pushing and holding the mute and mode buttons together for 1 second. The humidifier will beep when the value has been correctly stored.

9.Exit the voltage calibration test by pushing and holding down the mute button for longer than one second.

8.1.5Humidifier Warm-up and Control Check

This section describes how to check the humidifier's heater control systems. This test should be performed if there is a problem with the humidifier, or after servicing of the humidifier.

Equipment Required:

MR850 Humidifier

MR850 Heater Wire Adaptor

Flow/Temperature Probe

Suitable breathing circuit for MR850 (for example: RT100)

Suitable chamber for MR850 (for example: MR290) filled with water

Gas supply - constant flow of: 10±5 SLPM (Standard Litres Per Minute)

1.Make sure that the humidifier passes the calibration and display checks (refer 8.1.2 and 8.1.3).

2.Set up the humidifier as shown in section 4.1. Make sure the chamber probe is correctly inserted and invasive mode selected.

46 MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)