Symptom | Corrective Action | Reference | |
Heater wire alarm | Section 4.1.2 | ||
not working | wire circuit or disable this mode via the diagnostic menu. |
| |
Airway Probe | 1. | Check that the airway probe is inserted into the breathing |
|
alarm flashes along |
| circuit correctly, the breathing circuit assembled correctly, |
|
with an audible |
| and that there is water in the chamber. |
|
alarm | 2. | Check that the circuit is connected correctly to the |
|
|
| ventilator – gas flow could be reversed through the |
|
|
| humidifier. |
|
| 3. | Check that the gas flow rate is within specification of the | Section 3.4 |
|
| humidifier and type of accessories being used. Adjust as |
|
|
| necessary. |
|
| 4. | Check for excessive condensate build up. Excessively |
|
|
| cold or drafty ambient conditions may cause this alarm to |
|
|
| occur. Ensure there are no strong drafts around the |
|
|
| breathing circuit. |
|
| 5. | Complete a probe accuracy check. Replace probe as | Section 8.2 |
|
| necessary. |
|
| 6. | Humidifier faulty. Replace PCBs. |
|
Temperature / Flow | 1. | If the Temperature / Flow Probe alarm occurs with | Section 4.4 |
Probe alarm with |
| chamber or airway indicators also flashing, the |
|
airway or chamber |
| temperature probe is faulty. |
|
indicators flashing | 2. | Complete a probe accuracy check, and replace probe if | Section 8.2 |
|
| required. |
|
Temperature probe | 1. | Is the probe connector correctly plugged into the |
|
connector indicator |
| humidifier? Plug in probe as required. |
|
flashes, | 2. | Check that the circuit is connected correctly to the |
|
accompanied by an |
| ventilator – gas flow could be reversed through the |
|
audible alarm |
| humidifier. |
|
| 3. | Perform a humidifier calibration check. | Section 8.1.2 |
| 4. | Replace temperature probe and | Section 8.2 |
|
| condition disappears, the temperature probe is faulty. |
|
|
| Discard faulty probe. |
|
| 5. | Humidifier faulty. Replace PCBs. | Section 7.2.3 |
Temperature | The humidifier has been unable to maintain temperature over a |
| |
Indicator flashes, | period of time. |
| |
with audible alarm, | 1. | Gas flow has been disconnected from the humidifier either |
|
coupled with a low |
| reconnect gas flow or turn the humidifier off. |
|
temperature | 2. | Check that the gas flow rate is within specification of the |
|
(< 35.5 °C) |
| ||
| humidifier and accessories being used. |
| |
displayed |
|
| |
3. | Check for drafts around the breathing circuit. This can be |
| |
|
| ||
|
| caused by fans or room air conditioning. If this is found to |
|
|
| be the cause, the breathing circuit should be shielded from |
|
|
| the ambient airflow. |
|
| 4. | Check that the circuit is connected correctly to the |
|
|
| ventilator – gas flow could be reversed through the |
|
|
| humidifier. |
|
WARNING: Never cover the breathing circuit.
5. | Check for excessive condensate pooling in the breathing |
|
| circuit. Drain circuit if necessary. |
|
6. | Humidifier or probe faulty? Complete humidifier & probe | Section 8 |
| performance test. Replace probe or service humidifier as |
|
| required. |
|
7.Check that there is sufficient water in the chamber. Refill as necessary. Check that the water bag and delivery tube are not occluded.
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 25 |