Symptom

Corrective Action

Reference

Heater wire alarm

Non-heater wire mode has been activated, connect a heated

Section 4.1.2

not working

wire circuit or disable this mode via the diagnostic menu.

 

Airway Probe

1.

Check that the airway probe is inserted into the breathing

 

alarm flashes along

 

circuit correctly, the breathing circuit assembled correctly,

 

with an audible

 

and that there is water in the chamber.

 

alarm

2.

Check that the circuit is connected correctly to the

 

 

 

ventilator – gas flow could be reversed through the

 

 

 

humidifier.

 

 

3.

Check that the gas flow rate is within specification of the

Section 3.4

 

 

humidifier and type of accessories being used. Adjust as

 

 

 

necessary.

 

 

4.

Check for excessive condensate build up. Excessively

 

 

 

cold or drafty ambient conditions may cause this alarm to

 

 

 

occur. Ensure there are no strong drafts around the

 

 

 

breathing circuit.

 

 

5.

Complete a probe accuracy check. Replace probe as

Section 8.2

 

 

necessary.

 

 

6.

Humidifier faulty. Replace PCBs.

 

Temperature / Flow

1.

If the Temperature / Flow Probe alarm occurs with

Section 4.4

Probe alarm with

 

chamber or airway indicators also flashing, the

 

airway or chamber

 

temperature probe is faulty.

 

indicators flashing

2.

Complete a probe accuracy check, and replace probe if

Section 8.2

 

 

required.

 

Temperature probe

1.

Is the probe connector correctly plugged into the

 

connector indicator

 

humidifier? Plug in probe as required.

 

flashes,

2.

Check that the circuit is connected correctly to the

 

accompanied by an

 

ventilator – gas flow could be reversed through the

 

audible alarm

 

humidifier.

 

 

3.

Perform a humidifier calibration check.

Section 8.1.2

 

4.

Replace temperature probe and re-test. If the alarm

Section 8.2

 

 

condition disappears, the temperature probe is faulty.

 

 

 

Discard faulty probe.

 

 

5.

Humidifier faulty. Replace PCBs.

Section 7.2.3

Temperature

The humidifier has been unable to maintain temperature over a

 

Indicator flashes,

period of time.

 

with audible alarm,

1.

Gas flow has been disconnected from the humidifier either

 

coupled with a low

 

reconnect gas flow or turn the humidifier off.

 

temperature

2.

Check that the gas flow rate is within specification of the

 

(< 35.5 °C)

 

 

humidifier and accessories being used.

 

displayed

 

 

3.

Check for drafts around the breathing circuit. This can be

 

 

 

 

 

caused by fans or room air conditioning. If this is found to

 

 

 

be the cause, the breathing circuit should be shielded from

 

 

 

the ambient airflow.

 

 

4.

Check that the circuit is connected correctly to the

 

 

 

ventilator – gas flow could be reversed through the

 

 

 

humidifier.

 

WARNING: Never cover the breathing circuit.

5.

Check for excessive condensate pooling in the breathing

 

 

circuit. Drain circuit if necessary.

 

6.

Humidifier or probe faulty? Complete humidifier & probe

Section 8

 

performance test. Replace probe or service humidifier as

 

 

required.

 

7.Check that there is sufficient water in the chamber. Refill as necessary. Check that the water bag and delivery tube are not occluded.

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)

25