Table 7-1. 5725A Fault Codes

 

 

 

 

 

 

FAULT

 

 

MESSAGE

 

EXPLANATION

NO.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

201

5725

Self-Test ROM Failure

 

Requires service

202

5725

Self-Test RAM Failure

 

Requires service

 

203

5725

Self-Test EEPROM Failure

 

Requires service

 

204

5725

Self-Test Data Bus Failure

 

Requires service

 

205

5725

Self-Test CLAMPS Circuit Failure

 

Requires service

 

206

5725

Self-Test HVCLR Circuit Failure

 

Requires service

 

207

5725

Self-Test DAC Failure

 

Requires service

 

208

5725

Self-Test Watchdog Timer Failure

 

Requires service

 

209

5725

Current Heatsink Too Hot

 

Requires service

 

210

 

Output Tripped To Standby

 

Current heat sink too hot in operate

 

 

211

5725

Current Compliance Voltage Too

 

Internal fault, requires service

 

 

 

High

 

 

 

212

5725

Current Compliance Voltage Too

 

Most likely an external fault

 

 

 

High

 

 

 

213

5725

+400V Supply Did Not Shut Off

 

Requires service

 

214

5725

-400V Supply Did Not Shut Off

 

Requires service

 

215

5725

Voltage Heatsink Too Hot

 

Requires service

 

216

5725

Voltage Heatsink Too Hot

 

Voltage heat sink too hot in operate

 

217

5725

+400V Supply Too Small

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

218

5725

+400V Supply Too Large

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

219

5725

-400V Supply Too Large

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

220

5725

-400V Supply Too Small

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

221

5725

+400V Supply Current Too High

 

Internal fault, requires service

 

222

 

Output Tripped To Standby

 

+400V supply too high in operate, most likely an

 

 

 

 

 

 

 

external fault

223

5725

-400V Supply Current Too High

 

Internal fault, requires service

 

224

 

Output Tripped To Standby

 

-400V supply magnitude too high in operate,

 

 

 

 

 

 

 

most likely an external fault

225

5725

Fan Not Working

 

Requires service

 

226

5725

CLAMPS Fault

 

Internal fault, requires service

 

227

 

Output Tripped To Standby

 

CLAMPS circuit activated most likely because of

 

 

 

 

 

 

 

external transient condition

 

 

 

 

 

 

7-4