5725A

Instruction Manual

Table 1-19. 5700A Series II AC Current Specifications: 95 % Confidence Level

 

5700A

 

 

 

 

 

 

 

95%

 

95 % Confidence Level

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Absolute Uncertainty

 

Range

Resolution

Frequency

 

±5 °C from calibration temperature

 

 

 

 

 

 

For fields strengths >0.4 V/m but 3 V/m,

 

 

 

 

 

 

 

add 1 % of range.

Relative Uncertainty

±1 °C

Hz

24 Hours

90 Days

180 Days

1 Year

±(ppm output + nA)

24 Hours

90 Days

±(ppm output + nA)

220μA

2.2 mA

22 mA

1 nA

10 nA

100 nA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

550 + 25

280 + 20

100 + 16

400 + 40

1300 + 80

550 + 40

280 + 35

100 + 35

400 + 400

1300 + 800

550 + 400

280 + 350

100 + 350

400 + 4000

1300 + 8000

600 + 25

310 + 20

120 + 16

500 + 40

1400 + 80

600 + 40

310 + 35

120 + 35

500 + 400

1400 + 800

600 + 400

310 + 350

120 + 350

500 + 4000

1400 + 8000

650 + 25

330 + 20

130 + 16

550 + 40

1500 + 80

650 + 40

330 + 35

130 + 35

550 + 400

1500 + 800

650 + 400

330 + 350

130 + 350

550 + 4000

1500 + 8000

700 + 25

350 + 20

140 + 16

600 + 40

1600 + 80

700 + 40

350 + 35

140 + 35

600 + 400

1600 + 800

700 + 400

350 + 350

140 + 350

600 + 4000

1600 + 8000

375 + 25

220 + 20

90 + 16

375 + 40

1200 + 80

375 + 40

220 + 35

90 + 35

375 + 400

1200 + 800

375 + 400

220 + 350

90 + 350

375 + 4000

1200 + 8000

400 + 25

250 + 20

100 + 16

400 + 40

1200 +80

400 + 40

250 + 35

100 + 35

400 + 400

1200 + 800

400 + 400

250 + 350

100 + 350

400 + 4000

1200 + 8000

Hz

±(ppm output + μA)

±(ppm output + μA)

220 mA

2.2 A

1μA

10 μA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

020 - 1 k

01 k - 5 k

05 k - 10 k

550 + 4

280 + 3.5

100 + 3.5

400 + 40

1300 + 80

500 + 35

600 + 80

6500 + 160

600 + 4

310 + 3.5

120 + 3.5

500 + 40

1400 + 80

550 + 35

650 + 80

7500 + 160

650 + 4

330 + 3.5

130 + 3.5

550 + 40

1500 + 80

600 + 35

700 + 80

8000 + 1600

700 + 4

350 + 3.5

140 + 3.5

600 + 40

1600 + 80

650 + 35

750 + 80

8500 + 160

375 + 4

220 + 3.5

90 + 3.5

375 + 40

1200 + 80

500 + 35

550 + 80

6000 + 160

400 + 4

250 + 3.5

100 + 3.5

400 + 40

1200 + 80

550 + 35

650 + 80

7000 + 160

11 A

100μA

040 - 1 k

01 k - 5 k

05 k - 10 k

5725A Amplifier:

370

+ 170

400

+ 170

440

+ 170

460

+ 170

800

+ 380

850

+ 380

900

+ 380

950

+ 380

3000

+ 750

3300

+ 750

3500

+ 750

3600

+ 750

300 + 170

700 + 380

2800 + 750

330 + 170

800 + 380

3200 + 750

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 μA and

2.2mA ranges are increased by a factor of 1.3 plus 2 μA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.

1-26

Page 42
Image 42
Fluke 5725A instruction manual Relative Uncertainty ±1 C Hours Days Year ± ppm output + nA

5725A specifications

The Fluke 5725A is a high-performance, multifunctional temperature calibration source designed to meet the demanding requirements of laboratory and industrial environments. Renowned for its accuracy, reliability, and versatility, the 5725A serves as a powerful tool for engineers and technicians engaged in temperature calibration processes.

At the core of the 5725A is its advanced measurement technology, which allows for precision temperature calibrations across a wide range. With an operating temperature range from -200°C to 660°C, it caters to various applications, making it suitable for calibrating thermocouples, resistance temperature detectors (RTDs), and other temperature measurement devices.

One of the standout features of the Fluke 5725A is its outstanding accuracy, boasting a specified uncertainty of just ±0.15°C. This level of precision ensures that users can maintain compliance with regulatory standards and regulatory requirements in fields such as aerospace, pharmaceuticals, and manufacturing.

The Fluke 5725A is equipped with dual-channel capabilities, allowing users to calibrate two devices simultaneously. This enhances productivity by minimizing downtime and optimizing workflow efficiency. The intuitive touchscreen interface simplifies operations, enabling users to view measurements, set up test parameters, and access historical data with ease.

Versatile connectivity options such as USB and Ethernet enable seamless integration into existing systems and allow for data transfer, remote control, and configuration. This feature empowers technicians to conduct calibrations efficiently, whether on-site or remotely.

The robust design of the Fluke 5725A ensures its durability in demanding environments. Built with high-quality materials, it is resistant to shocks and vibrations, making it suitable for use in manufacturing plants and laboratories where equipment may be subjected to harsh conditions.

Another notable characteristic of the 5725A is its ability to generate stable temperature outputs with minimal drift, contributing to consistency in calibration results. Its built-in self-test functionality helps ensure that the unit remains in peak operating condition, giving users confidence in their calibration processes.

In summary, the Fluke 5725A is a top-tier temperature calibration source that excels in accuracy, reliability, and user-friendliness. Whether in laboratory settings or industrial applications, its advanced technologies and features make it an essential tool for professionals tasked with maintaining precision in temperature measurements.