Introduction and Specifications 1

AC Current Specifications

1-16. AC Current Specifications

5720A

99%

Range

Table 1-16. 5720A Series II AC Current Specifications: 99 % Confidence Level

99 % Confidence Level

 

 

 

Absolute Uncertainty

 

Relative Uncertainty

Resolution

Frequency

±5 °C from calibration temperature

 

±1 °C

 

 

For fields strengths >0.4 V/m but 3 V/m,

 

 

 

 

 

add 1 % of range.

 

 

 

 

 

 

 

 

Hz

24 Hours

90 Days

180 Days

±(ppm output + nA)

1 Year

24 Hours

90 Days

±(ppm output + nA)

220μA

2.2mA

22 mA

1nA

10 nA

100 nA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

5k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

5k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

5k - 10 k

260 + 20

170 + 12

120 + 10

300 + 15

1000 + 80

260 + 50

170 + 40

120 + 40

210 + 130

1000 + 800

260 + 500

170 + 400

120 + 400

210 + 700

1000 + 6000

280 + 20

180 + 12

130 + 10

320 + 15

1100 + 80

280 + 50

180 + 40

130 + 40

220 + 130

1100 + 800

280 + 500

180 + 400

130 + 400

220 + 700

1100 + 6000

290 + 20

190 + 12

135 + 10

340 + 15

1200 + 80

290 + 50

190 + 40

135 + 40

230 + 130

1200 + 800

290 + 500

190 + 400

135 + 400

230 + 700

1200 + 6000

300 + 20

200 + 12

140 + 10

350 + 15

1300 + 80

300 + 50

200 + 40

140 + 40

240 + 130

1300 + 800

300 + 500

200 + 400

140 + 400

240 + 700

1300 + 6000

260 + 20

130 + 12

100 + 10

250 + 15

900 + 80

260 + 50

130 + 40

100 + 40

250 + 130

900 + 800

260 + 500

130 + 400

100 + 400

250 + 700

900 + 6000

280 + 20

150 + 12

110 + 10

280 + 15

1000 + 80

280 + 50

150 + 40

110 + 40

280 + 130

1000 + 800

280 + 500

150 + 400

110 + 400

280 + 700

1000 + 6000

Hz

±(ppm output + μA)

±(ppm output + μA)

220 mA

2.2 A

1μA

10 μA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

5k - 10 k

20 - 1 k

1 k - 5 k

5 k - 10 k

260 + 5

170 + 4

120 + 3

210 + 4

1000 + 12

290 + 40

440 + 100

6000 + 200

280 + 5

180 + 4

130 + 3

220 + 4

1100 + 12

300 + 40

460 + 100

7000 + 200

290 + 5

190 + 4

135 + 3

230 + 4

1200 + 12

310 + 40

480 + 100

7500 + 200

300 + 5

200 + 4

140 + 3

240 + 4

1300 + 12

320 + 40

500 + 100

8000 + 200

260 + 5

130 + 4

100 + 3

250 + 4

900 + 12

300 + 40

500 + 100

6000 + 200

280 + 5

150 + 4

110 + 3

280 + 4

1000 + 12

350 + 40

520 + 100

7000 + 200

11 A

100μA

5725A Amplifier:

40 - 1 k

370

+ 170

400

+ 170

440

+ 170

460

+ 170

1 k - 5 k

800

+ 380

850

+ 380

900

+ 380

950

+ 380

5 k - 10 k

3000

+ 750

3300

+ 750

3500

+ 750

3600

+ 750

 

 

 

 

 

 

 

 

 

300

+ 170

330

+ 170

700

+ 380

800

+ 380

2800

+ 750

3200

+ 750

 

 

 

 

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 μA and 2.2 mA ranges are increased by a factor of 1.3 plus 2 μA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.

1-23

Page 39
Image 39
Fluke 5725A instruction manual AC Current Specifications, Absolute Uncertainty Relative Uncertainty Resolution

5725A specifications

The Fluke 5725A is a high-performance, multifunctional temperature calibration source designed to meet the demanding requirements of laboratory and industrial environments. Renowned for its accuracy, reliability, and versatility, the 5725A serves as a powerful tool for engineers and technicians engaged in temperature calibration processes.

At the core of the 5725A is its advanced measurement technology, which allows for precision temperature calibrations across a wide range. With an operating temperature range from -200°C to 660°C, it caters to various applications, making it suitable for calibrating thermocouples, resistance temperature detectors (RTDs), and other temperature measurement devices.

One of the standout features of the Fluke 5725A is its outstanding accuracy, boasting a specified uncertainty of just ±0.15°C. This level of precision ensures that users can maintain compliance with regulatory standards and regulatory requirements in fields such as aerospace, pharmaceuticals, and manufacturing.

The Fluke 5725A is equipped with dual-channel capabilities, allowing users to calibrate two devices simultaneously. This enhances productivity by minimizing downtime and optimizing workflow efficiency. The intuitive touchscreen interface simplifies operations, enabling users to view measurements, set up test parameters, and access historical data with ease.

Versatile connectivity options such as USB and Ethernet enable seamless integration into existing systems and allow for data transfer, remote control, and configuration. This feature empowers technicians to conduct calibrations efficiently, whether on-site or remotely.

The robust design of the Fluke 5725A ensures its durability in demanding environments. Built with high-quality materials, it is resistant to shocks and vibrations, making it suitable for use in manufacturing plants and laboratories where equipment may be subjected to harsh conditions.

Another notable characteristic of the 5725A is its ability to generate stable temperature outputs with minimal drift, contributing to consistency in calibration results. Its built-in self-test functionality helps ensure that the unit remains in peak operating condition, giving users confidence in their calibration processes.

In summary, the Fluke 5725A is a top-tier temperature calibration source that excels in accuracy, reliability, and user-friendliness. Whether in laboratory settings or industrial applications, its advanced technologies and features make it an essential tool for professionals tasked with maintaining precision in temperature measurements.