Fluke 5725A instruction manual Selecting Line Voltage

Models: 5725A

1 208
Download 208 pages 33.86 Kb
Page 53
Image 53
Selecting Line Voltage

Installation 2

Selecting Line Voltage

2-7. Selecting Line Voltage

The 5725A arrives from the factory configured for the line voltage normally appropriate for the country of purchase, or as specified at the time of purchase. Check the line power label on the rear panel of the 5725A to verify that the line voltage matches local line power. Figure 2-3 shows the location of the line power label and switches, and the switch settings for each line voltage.

The 5725A can be set to operate from eight different nominal line voltages; each voltage setting has a voltage tolerance of ±10% and a frequency range of 47 to 63 Hz. The line voltage switches are located on the bottom left side of the rear panel.

To change the line voltage setting, set the line voltage selection switches to the correct setting shown in Figure 2-3.

CAUTION FOR FIRE PROTECTION

REPLACE ONLY WITH A 250V FUSE

OF INDICATED RATING.

VOLTAGE

FUSE

SELECTION

 

S2

S3

S4

100V

 

!

110V

 

F8A 250V

115V

 

(FAST)

120V

 

 

S2 S3 S4

200V 2-7. Selecting Line Voltage 220V Manual background 230V Manual background 240V Manual background

!

F4A 250V

(FAST)

WARNING GROUNDING

CONNECTOR IN POWER CORD

MUST BE CONNECTED TO

ENSURE PROTECTION FROM

ELECTRONIC SHOCK.

47-63 Hz 750VA MAX

!

FUSE

CHASSIS

GROUND

S2

CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING.

VOLTAGE

FUSE

SELECTION

S2

S3

S4

100V

 

!

110V

 

F8A 250V

115V

 

(FAST)

120V

 

 

S2

S3

S4

200V

 

!

220V

 

F4A 250V

230V

 

(FAST)

240V

 

 

S3

 

S4

JOHN FLUKE MFG. CO., INC.

MADE IN U.S.A.

PATENTS PENDING

CURRENT

NO INTERNAL USER SERVICEABLE

PARTS REFER SERVICE TOOUTPUT QUALIFIED SERVICE PERSONNEL

TO CLEAN FILTER

REMOVE FROM INSTRUMENTHI AND FLUSH WITH WARM

SOAPY WATER

11A MAX

20V PK

LO

MAX

 

5700A CALIBRATOR

CALIBRATION

ENABLE NORMAL

aq4f.eps

Figure 2-3. Line Power Label and Switch Location

2-7

Page 53
Image 53