1-17. Wideband AC Voltage (Option 5700-03) Specifications

Table 1-21. Wideband AC Voltage (Option 5700-03) Specifications

Specifications apply to the end of the cable and 50 Ω termination used for calibration

Range

Resolution

Absolute Uncertainty

±5 °C from calibration temperature

30 Hz-500 kHz

Volts

dBm

24 Hours

90 Days

180 Days

±(% output + μV)

1 Year

1.1mV

3 mV

11 mV

33 mV

110 mV

330 mV

1.1V

3.5V

-46

-37

-26

-17

-6.2

+3.4

+14

+24

10nV

10nV

100nV

100nV

1μV

1μV

10μV

10μV

0.4+ 0.4

0.4+ 1

0.2+ 4

0.2+ 10

0.2+ 40

0.2+ 100

0.2+ 400

0.15+ 500

0.5+ 0.4

0.45+ 1

0.35+ 4

0.3+ 10

0.3+ 40

2.5+ 100

0.25+ 400

0.2+ 500

0.6+ 0.4

0.5+ 1

0.5+ 4

0.45+ 10

0.45+ 40

0.35+ 100

0.35+ 400

0.3+ 500

0.8+ 2

0.7+ 3

0.7+ 8

0.6+ 16

0.6+ 40

0.5+ 100

0.5+ 400

0.4+ 500

Frequency

Frequency

Amplitude Flatness, 1 kHz Reference

Temperature

 

Resolution

 

Voltage Range

 

Coefficient

 

 

 

 

 

 

 

 

1.1 mV

3 mV

>3 mV

 

Settling

Time To

Full

Accuracy

Harmonic Distortion

Hz

10 - 30

30 - 120

120 - 1.2 k

1.2k - 12 k 12 k - 120 k

120 k - 1.2 M

1.2M - 2 M

2 M - 10 M

10 M - 20 M

20 M - 30 M

Hz

0.01

0.01

0.1

1

10

100

100k

100k

1M

1M

±(% output + floor indicated)

 

0.3

 

 

0.3

 

 

0.3

 

0.1

 

 

0.1

 

 

0.1

 

 

 

 

 

 

0.1

 

 

0.1

 

 

0.1

 

 

 

 

 

 

0.1

 

 

0.1

 

 

0.1

 

0.1

 

 

0.1

 

 

0.1

0.2

+ 3 μV

 

0.1

+ 3 μV

 

0.1

+ 3 μV

0.2

+ 3 μV

0.1

+ 3 μV

0.1

+ 3 μV

0.4

+ 3 μV

 

 

+ 3 μV

 

 

+ 3 μV

0.3

0.2

0.6

+ 3 μV

 

 

+ 3 μV

 

 

+ 3 μV

 

0.5

 

0.4

10.5

+ 15 μV

10.5

+ 3 μV

1

+ 3 μV

±ppm/°C

100

100

100

100

100

100

100

100

150

300

Seconds

7

7

5

5

5

5

0.5

0.5

0.5

0.5

dB

-40 -40 -40 -40 -40 -40 -40 -40 -34 -34

Additional Operating Information:

dBm reference = 50 Ω

Range boundaries are at voltage points, dBm levels are approximate.

( Power )

dBm = 10 log; 0.22361 V across 50 Ω = 1 mW or 0 dBm 1 mW

Minimum output: 300 μV (-57 dBm)

Frequency uncertainty: ±0.01 %

Frequency resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz.

Overload protection: A short circuit on the wideband output will not result in damage. After settling time, normal operation is restored upon removal.