3-8.

Voltage Standby

 

 

 

 

 

NO OUTPUT

 

 

 

 

 

BOOST

 

 

 

 

 

LIT

 

 

 

 

 

 

 

 

 

 

 

OUTPUT

SENSE

 

 

CURRENT

 

 

 

V Ω A

V Ω A

 

μ

OUTPUT

 

VOLTAGE MODE

NO

HI

 

WIDEBAND

 

 

CURRENT MODE

OUTPUT

 

HI

 

HI

 

 

 

 

 

 

5725A STANDBY

 

 

 

 

 

 

 

 

 

 

 

 

11A

 

 

LO

 

LO

 

 

MAX

FAULT

 

 

 

LO

 

 

 

 

 

 

 

OVERLOAD

 

 

 

 

 

 

2OV

 

 

 

 

 

 

PK

CABLE OFF

 

HI

 

 

 

 

MAX

 

 

 

 

 

 

 

 

AUX

GUARD

GROUND

 

 

 

 

CURRENT

 

 

 

 

 

 

 

 

 

aq9i.eps

Figure 3-4. Voltage Standby Mode

Voltage standby mode is when both VOLTAGE MODE and 5725A STANDBY indicators are lit. In voltage standby mode, the 5725A is not supplying an output. Voltage standby occurs if "5725A" is selected in a 5700A setup menu as the voltage amplifier, and the next two conditions are true:

An ac voltage 220V or above is showing on the 5700A Output Display with the STANDBY annunciator lit.

The BOOST indicator on the 5700A is lit.

Note

If condition 1 is true, the BOOST indicator can be toggled on and off by pressing the 5700A BOOST key. If the BOOST indicator is off, the 5725A is in standby mode, not voltage standby.