Page 26
Image 26
Mode 1

Section 4

5.To exit the manual mode, press CLEAR ENTRY. This returns the sampler to the standby state.

While the sampler is in standby, a sample cycle can be initiated by pressing the TAKE SAMPLE key.

4.8 Measurements Triggered by Liquid Levels or Flow

The sampler can operate in any one of three control modes, using an external controlling device (i.e. Liquid Level Actuator, Cat. No. 943, or an external flow meter). In any of the three modes, the sampler remains in a standby state until liquid at the sample source reaches a predetermined level. This feature may be also used to access the setup of the storm water monitoring program, First Flush. The three modes are described below:

Mode 1

The instrument collects a sample within the first minute that the liquid rises to the predetermined level. After the first sample is taken, the sample interval automatically resets to the programmed interval.

If the liquid falls below the predetermined level, the sampler continues to collect samples at the programmed time or flow interval, until the program ends.

Mode 2

The instrument collects a sample within the first minute that liquid rises to the predetermined level. After the first sample is taken, the sample interval automatically resets to the programmed interval. The sample program ends if liquid falls below the predetermined level. If the liquid again rises to the predetermined level, a sample is taken within the first minute and the interval automatically resets to the programmed time or flow interval.

Mode 3

When the liquid rises to a predetermined level, a sample is not taken immediately. A sample is taken only when liquid is at or above the predetermined level and after the programmed sample interval has elapsed. The interval between samples counts down and resets even when the liquid level is below the predetermined (actuation) level. This mode is used to synchronize sample collection with a real-time clock when sampling in the time mode.

Page 26

 

Measurements Triggered by Liquid Levels or Flow

8544operations.fm

Page 26
Image 26
Hach 900 manual Section, Mode

FAQ

What is the function of the CLEAR ENTRY key?

The CLEAR ENTRY key is used to exit the manual mode and return the sampler to the standby state.

Can the sampler operate using an external controlling device?

Yes, the sampler can operate using an external controlling device such as a Liquid Level Actuator or an external flow meter.

What are the three control modes of the sampler?

The three control modes are Mode 1, Mode 2, and Mode 3.

How to exit the manual mode?

To exit the manual mode, press CLEAR ENTRY.

How does the sampler operate in Mode 1?

The instrument collects a sample within the first minute that the liquid rises to the predetermined level and resets the sample interval to the programmed interval.

How does the sampler operate in Mode 2?

The instrument collects a sample within the first minute that liquid rises to the predetermined level and resets the sample interval to the programmed interval. The sample program ends if liquid falls below the predetermined level.

How does the sampler operate in Mode 3?

The sampler takes a sample only when liquid is at or above the predetermined level and after the programmed sample interval has elapsed.

What is the purpose of the Liquid Level Actuator?

The Liquid Level Actuator is used as an external controlling device for the sampler.

How to access the setup of the storm water monitoring program?

The setup of the storm water monitoring program can be accessed using the Liquid Level Actuator or an external flow meter.

What is the function of the TAKE SAMPLE key?

While the sampler is in standby, a sample cycle can be initiated by pressing the TAKE SAMPLE key.