•Access the Sys Par 2 menu as shown on page 21.
SYS PAR 1 | 09:07:31 |
|
|
|
CLOCK TEMPS IN °F |
|
|
| |
PROBES | LOGGING |
|
|
|
SETUP MENU | MORE | SYS PAR 2 | 09:08:41 | |
|
| REMOTE ALARM MODE | ||
|
| ALARMS |
| BUZZER |
|
| BACK | FAC PRESETS |
REMOTE ALARM
ACTIVATE REMOTE
ALARM WHEN: PRINTER ERROR =NO
| ENTER |
ACTIVATE REMOTE | |
ALARM WHEN: |
|
CYCLE ENDS | =NO |
| ENTER |
ACTIVATE REMOTE | |
ALARM WHEN: |
|
DOOR OPEN | =NO |
| ENTER |
ACTIVATE REMOTE
ALARM WHEN:
HIGH/LOW TEMPS=YES
ENTER
ACTIVATE REMOTE
ALARM WHEN:
POWER FAILURE =YES
ENTER
ACTIVATE REMOTE ALARM WHEN:
CYCLE RUNNING =NO ENTER
•From Sys Par 2, select REMOTE ALARM.
•Yes closes the Remote Alarm circuit in the event of a printer error. (Use to change No to Yes, etc.)
•Press to move to the next field.
•Yes closes the Remote Alarm circuit when the cycle ends. (Use to change No to Yes, etc.)
•Press to move to the next field.
•Yes closes the Remote Alarm circuit whenever the door is open for longer than the ‘Door Open’ alarm time setting on page 25. (Use to change No to Yes, etc.)
•Press to move to the next field.
•Yes closes the Remote Alarm circuit if temperature is sensed above the High Alarm Temperature limit or below the Low Alarm Temperature limit. (Use to change Yes to No, etc.) Refer to pages 26 and 28 to set the High Alarm and Low Alarm temperatures.
•Press to move to the next field.
•Yes closes the Remote Alarm circuit when power is restored after a power interruption. (Use to change Yes to No, etc.)
•Press to move to the next field.
•Yes closes the Remote Alarm circuit when a cycle is running. The Cycle Running option has a higher precedence and overrides all other remote alarm options, resetting them to NO. (Use to change Yes to No, etc.)
(Pressing returns to . . . Printer Error.)
•Select ENTER to keep any changes and return to SYS PAR 2.
– 24 –