Figure 5-4. Low Frequency Response Test

AC VOLTMETER ACCURACY CHECK.

a. A

Voltmeter

Calibrator

Model 738BR)

will

be required

for this

operation.

b. Adjust voltmeter calibrator for 400 cps- rms output.

c. Set Model 410C FUNCTION SELECTOR to ACV; RANGE to 500 V.

d. Adjust the voltmeter calibrator to settings listed in Table 5-6.Model 410C should indicate readings within limits specified. If not, refer

5-25. AC VOLTMETER FREQUENCY RESPONSE CHECK.

a. A Frequency Response Test Set Model

739AR), a Test Oscillator Model 651A), an

RF Signal Generator

Model 608 C), a Power

Meter

Model 431

B), a Thermistor

Mount

Model 478A), a Probe - T - Connector

Model 11042A), a VHF Signal Generator

Model 612A) and a 10

KC Filter

Model K02-

411A)

will be required for this operation. Fig-

ure 5-3

and 5-4 describe the arrangement

to be