TM 11-6625-1614-15

Section V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model 410C

Paragraphs 5-12to 5-15

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table

5-2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 5-2. DCV Accuracy Test

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model 410C

 

 

 

Voltmeter

Calibrator

 

 

 

 

 

 

Model 410C

 

 

 

Range

Settings

 

 

 

Model

738B

 

 

 

 

 

 

Meter

Readings

 

 

 

 

 

 

 

 

 

Settings

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Range

Voltage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

. 015V

 

 

 

1. 5-5

.015

 

 

 

 

 

.01.47 to .0153 V

 

 

. 0 5 V

 

 

 

1. 5-5

.05

 

 

 

 

 

 

.049

to .051

V

 

 

 

. 1 5 V

 

 

 

1. 5-5

.15

 

 

 

 

 

 

.147 to .153 V

 

 

 

. 5 V

 

 

 

 

1. 5-5

. 5

 

 

 

 

 

 

.49 to .51

V

 

 

 

 

1 . 5 V

 

 

 

1. 5-5

1.5

 

 

 

 

 

 

1.47

to

l.53

V

 

 

 

5 V

 

 

 

 

1. 5-5

5

 

 

 

 

 

 

4.9

to

5.l

V

 

 

 

 

15 V

 

 

 

 

1. 5-5

15

 

 

 

 

 

 

14.7

to

15.3

V

 

 

 

50 V

 

 

 

 

1. 5-5

50

 

 

 

 

 

 

49 to 51 V

 

 

 

 

 

150V

 

 

 

1.5-5

150

 

 

 

 

 

147 to 153 V

 

 

 

 

500 V

 

 

 

1-3

300

 

 

 

 

 

290 to 310 V

 

 

 

 

1500

 

 

 

 

1-3

300

 

 

 

 

 

 

270 to 330 V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

b. Adjust

Voltmeter

Calibrator to provide a

5-14.

ACCURACY

CHECK

(DCA).

 

 

 

 

+.015 v

dc

voltage.

 

 

 

 

 

a. Figure 5-2describes the test arrangement

 

 

 

 

 

 

 

 

 

 

c. Model

410C should

read

between

0. 0147 and

 

required

for

this operation.

The

following

 

 

additional

equipment will

also

be

required:

 

0.0153 v.

 

 

 

 

 

 

 

 

 

 

 

 

DC Power Supply

Model 723A)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

d. Readjust Model 410C

and Voltmeter Cali-

DC Voltmeter

 

 

Model 3440A/3442A)

 

 

brator (,) settings listed in Table 5-2. Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model 410C meter readings. If Model 410C fails

10

K,

1%,

 

1

w

resistor

 

Part No.

0727-0157)

 

to meet

pecifications,

refer to Paragraph, 5-30

56

K,

1%,

 

1

w resistor

Part

No.

0730-0053)

 

and 5−32

for proper

adjustment

procedure .

 

 

10

, 1%, 1

w

resistor

Part

No.

0727-0335)

5-12

 

 

 

 

 

 

 

INPUT RESISTANCE CHECK (DCV).

56 ,

1%,

1/2

w

resistor

Part

No. 0811-0341)

 

 

 

 

 

 

 

 

 

a. Connect an external resistor, Rx, of lO M

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ohms ±1% ( Part No. 0370-0168) in series

 

b. Connect

the

Model

410C

as shown

in Figure

 

between the voltmeter calibrator and the DCV

 

5-2; FUNCTION SELECTOR to +DCA; RANGE

 

cable of the Model 410C.

 

 

 

 

to 150 MA.

 

 

 

 

 

 

 

 

 

b. Set Model 410C FUNCTION selector to +DCV;

 

c. Use 56 ohm resistor for R1 and 10 ohm

 

RANGE to .015 V.

 

 

 

 

 

resistor

for

R2.

 

 

 

 

 

 

 

c. Adjust voltmeter calibrator for +.015v DC

 

d. Adjust dc power supply to obtain 1.4v reading

 

output.

 

 

 

 

 

 

 

on

system

voltmeter.

 

 

 

 

 

 

 

d. Model 410C should read .0075

v, verifying

 

e. Model

410C

should read

between 135.5 and

 

Rin of 10 M ohms.

 

 

 

 

 

 

 

 

 

 

 

144.5 ma.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

e. Table 5-3provides settings required to

 

f. Adjust dc power supply for System voltmeter

 

verify Model 410C Rin on

RANGES specified.

 

readings listed in Table 5-4. Note Model 410C

 

 

 

 

 

 

 

 

 

meter readings.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5-15. INPUT RESISTANCE CHECK (DCA).

 

 

 

 

 

 

 

 

 

 

a. Figure 5-2describes the test arrangement

 

 

 

 

 

 

 

 

 

required

for

this operation.

 

Replace

R1 and

 

 

 

 

 

 

 

 

 

R2 with a 10 ohm ±1% resistor

Part No.

 

 

 

 

 

 

 

 

 

0727-0335).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

b. Set Model 410C FUNCTION SELECTOR to

5-13.

 

 

 

 

 

 

 

 

+DCA: RANGE to 150 MA.

 

 

 

 

5-2

C h a n g e 1

01556-2

Page 36
Image 36
HP 410C manual Input Resistance Check DCV, Input Resistance Check DCA