Language Dictionary - 4

*OPT?

This query requests the dc source to identify any options that are installed. Options are identified by number A 0 indicates no options are installed.

Query Syntax *OPT?

Returned Parameters <AARD>

*RCL

WARNING: Recalling a previously stored state may place hazardous voltages at the dc source output.

This command restores the dc source to a state that was previously stored in memory with the *SAV command to the specified location. All states are recalled with the following exceptions:

υthe trigger system is set to the Idle state by an implied ABORt command (this cancels any uncompleted trigger actions)

υthe calibration function is disabled by setting CAL:STATe to OFF

NOTE: The device state stored in location 0 is automatically recalled at power turn-on when the OUTPut:PON:STATe is set to RCL0.

Command Syntax

*RCL <NRf>

Parameters

0 1 2 3

Example

*RCL 3

Related Commands

*PSC *RST *SAV

*RST

This command resets the to a factory-defined state as defined in the following table. *RST also forces an ABORt command.

Command Syntax *RST

Parameters None

Related Commands *PSC *SAV