4 - Language Dictionary

CALibrate:STATe

This command enables and disables calibration mode. The calibration mode must be enabled before the will accept any other calibration commands.

The first parameter specifies the enabled or disabled state. The second parameter is the password. It is required if the calibration mode is being enabled and the existing password is not 0. If the password is not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query statement returns only the state, not the password.

NOTE: Whenever the calibration state is changed from enabled to disabled, any new calibration constants are lost unless they have been stored with CALibrate:SAVE.

Command Syntax

CALibrate:STATe<bool>[,<NRf>]

Parameters

0 1 OFF ON [,<password>]

*RST Value

OFF

Examples

CAL:STAT 1,6812 CAL:STAT OFF

Query Syntax

CALibrate:STATe?

Returned Parameters

<NR1>

Related Commands

CAL:PASS CAL:SAVE *RST

CALibrate:VOLTage

This command initiates the calibration of the output voltage and the voltage readback circuit.

Command Syntax

CALibrate:VOLTage[:DC]

Parameters

None

 

Examples

CAL:VOLT

CAL:VOLT:DC

CALibrate:VOLTage:PROTection

This command can calibrates the overvoltage protection (OV) circuit. The dc source automatically performs the calibration. CALibrate:VOLTage:PROTection is a sequential command that takes several seconds to complete.

Command Syntax

CALibrate:VOLTage:PROTection

Parameters

None

Examples

CAL:VOLT:PROT