Language Dictionary - 4

TRIGger:SEQuence2:LEVel:CURRent

TRIGger:ACQuire:LEVel:CURRent

Agilent 66312A, 66332A Only

This command sets the trigger level for internally triggered current measurements. A positive current trigger occurs when the current level changes from a value less than the lower hysteresis band limit to a value greater than the upper hysteresis band limit. Similarly, a negative current trigger occurs when the current level changes from a value greater than the upper hysteresis band limit to a value less than the lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:LEVel:CURRent<NRf+>

 

TRIGger:ACQuire:LEVel:CURRent<NRf+>

Parameters

0 to MAX (see table 4-3)

 

Unit

A (amperes)

 

*RST Value

0

 

Examples

TRIG:SEQ2:LEV:CURR 5

TRIG:ACQ:LEV:CURR MAX

 

TRIG:ACQ:LEV 2

 

Query Syntax

TRIGger:SEQuence2:LEVel:CURRent?

 

TRIGger:ACQuire:LEVel:CURRent?

Returned Parameters

<NR3>

 

Related Commands

TRIG:SEQ2:LEV:VOLT

TRIG:SEQ2:HYST:CURR

TRIGger:SEQuence2:LEVel:VOLTage

TRIGger:ACQuire:LEVel:VOLTage

Agilent 66312A, 66332A Only

This command sets the trigger level for internally triggered voltage measurements. A positive voltage trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the voltage level changes from a value greater than the upper hysteresis band limit to a value less than the lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:LEVel:VOLTage<NRf+>

 

TRIGger:ACQuire:LEVel:VOLTage<NRf+>

Parameters

0 to MAX (see table 4-3)

 

Unit

V (volts)

 

*RST Value

0

 

Examples

TRIG:SEQ2:LEV:VOLT 5

TRIG:ACQ:LEV:VOLT MAX

 

TRIG:ACQ:LEV 2

 

Query Syntax

TRIGger:SEQuence2:LEVel:VOLTage?

 

TRIGger:ACQuire:LEVel:VOLTage?

Returned Parameters

<NR3>

 

Related Commands

TRIG:SEQ2:LEV:CURR

TRIG:SEQ2:HYST:VOLT

77