Cintel International Ltd.

diTTo User Guide

Tools: Exposure Control

Warning.

Adjustments of the Exposure will invalidate the settings Of DMIN and may not comply with Cineon DPX standards.

Be careful with this adjustment, it is very easy to cause the scan to cut-off (clip) some of the image detail in the highlights or lowlights

Resetting DMIN will return Exposure values to a default.

The Exposure control provides the ability to individually adjust the ‘setting’ of the Red Green and Blue signals. The adjustment varies the light exposure time of the Red, Green and Blue illumination and can therefore make a minor change the scan time.

Multiple settings can be made and each setting can be named by Add to list

Named Exposure settings are saved with the current project

Fig 23 Exposure Control Window

Tools: Auto Shading Correction

Automatic shading compensation caters for any optical variations (RGB) and calibrates the camera cell white and black balances. It is customary to perform a shading align weekly only after verification of a clean optical path. (See maintenance section Page 26).

1.Perform only with an open gate. Unload servo and remove film. Ensure gate is in loaded position.(See page 21)

2.Select Auto Shading Correction

3.Follow the on-screen prompts (Note: operation takes approximately 5 Minutes)

4.The resultant Shading file is saved by default on the local C Drive and is automatically applied to all future scans.

Tools: DMIN Calibration (Minimum Density setting)

DMIN calibration gives automatic exposure balance to level 95 (~1% level) of the Cineon transfer function. This target level can be set to an alternate value in New Project Wizard (Page 12) or Project Settings (Page 31). See Appendix 1 for more details about DPX files and levels.

1.Load required film and transport to required clip

2.Select DMIN Calibration and follow the on screen prompts

The scanner will automatically rack the film until the framing bar is visible and then select a portion of the framing bar to use as the DMIN. The scanner will ask for confirmation at this point.

3.If the area automatically selected has some picture information visible or a better area is available, Use the mouse to drag a sample window over a new desired area. (Verification of minimum levels can be seen in the Pixel Picker) Select OK and wait until the calibration is complete.

Progress can be viewed on the pixel picker feedback at the bottom of the GUI by observing the RGB levels moving towards a level of 95 or other level if set differently in the Project Settings. Note that the values may not display as exactly 95 as the picker averages the pick area values including any dust etc. However the DMIN Calibrator ignores any values greater than the lowest and will therefore ignore dust etc. The film is now automatically calibrated to Cineon DMIN specification.

Fig 24 Pixel Picker Feedback

at bottom of GUI

Picker location and size

Average value

Average density

Issue 3