Quality Control Procedures

determine the film parameter values as follows:

a. Determine which step has a density closest to 2.20. Then determine which step has a density closest to but not less than 0.45. Designate the difference between these densities as Density Difference (DD).

b.Determine which step has an average density closest to but not less than 1.20. Designate this step as Medium Density (MD).

c. Designate the average for Step 1 (or the clear area on the film) as Base Plus Fog.

d.Record the numeric values of DD, MD and Base Plus Fog on the center lines of the laser imager QC Chart. See the Quality Control Chart on the next page. Record also the step numbers involved.

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

21

8600-11A

QC Step Wedge

January 10, 2006

5E6155

A-3

Page 117
Image 117
Kodak DryView 8900 manual Determine the film parameter values as follows

DryView 8900 specifications

The Kodak DryView 8900 is a state-of-the-art dry imaging system that revolutionizes medical imaging with its advanced technology and unparalleled performance. Designed primarily for the healthcare sector, this system delivers high-quality printed images for various diagnostic applications, making it an essential tool for hospitals, clinics, and imaging centers.

One of the standout features of the Kodak DryView 8900 is its exceptional image quality. The system utilizes advanced imaging algorithms and high-resolution printing capabilities to produce crisp, clear images that facilitate accurate diagnosis and enhanced patient care. This high fidelity is crucial in fields such as radiology, mammography, and other imaging modalities where precision is paramount.

Incorporating thermal imaging technology, the DryView 8900 eliminates the need for traditional wet processing chemicals, which is not only environmentally friendly but also simplifies the workflow in medical facilities. This dry imaging process results in faster turnaround times for print production, allowing clinicians quick access to essential imaging data.

The DryView 8900 also boasts a user-friendly interface, designed with healthcare professionals in mind. Its intuitive controls and straightforward operation minimize the learning curve, enabling users to navigate the system efficiently. Moreover, the printer supports a variety of film sizes and formats, providing flexibility to meet the diverse needs of medical imaging departments.

Another impressive characteristic of the Kodak DryView 8900 is its durability and reliability. Built with robust components, the system is designed to handle the rigorous demands of a busy clinical environment. Regular maintenance and servicing are streamlined, helping to maintain optimal performance and reduce downtime.

Additionally, the Kodak DryView 8900 incorporates connectivity options that allow seamless integration with existing imaging systems and PACS (Picture Archiving and Communication Systems). This connectivity ensures that the DryView 8900 can receive high-quality images directly from modalities, enhancing the overall efficiency of the imaging process.

In summary, the Kodak DryView 8900 is a cutting-edge imaging solution that combines remarkable image quality, user-friendly design, and reliable performance. With its dry imaging technology, broad compatibility, and environmental considerations, this system stands out as a vital asset in today’s medical imaging landscape, driving advancements in healthcare delivery and patient outcomes.