Kodak RP 30 manual Prerequisites for MBL and PBL Calibration, Printing a PBL Test, Test Start PBL

Models: RP 30

1 235
Download 235 pages 61.19 Kb
Page 136
Image 136

Operator’s Guide

Prerequisites for MBL and PBL Calibration

Before printing the MBL and PBL calibration:

1.Check that there are no print orders pending.

2.Use the Information (i) button to check that no paper is in the paper transport of the printer and the paper processor for the MBL and the PBL calibration.

3.Check that there is enough paper in the magazine for the two tests (approximately 3 m [40 in.]).

4.Check that the chemical solutions are at nominal temperature.

Printing a PBL Test

When it is necessary to print a PBL test, follow these instructions.

1.Touch:

Test

Start PBL

2.Select the paper configuration for which the PBL test is to be printed.

3.Touch PBL to print the test.

4.Enter in the next screen:

Emulsion no. (only for information, not assessed)

Tolerance: Step width 0.1 mm

Possible input of the manually measured deviation of the paper from the coded width (useful for border prints).

5.Activate / deactivate with Laser point.

6.Touch Start.

5-4

KODAK PROFESSIONAL RP 30 Laser Printer

Page 136
Image 136
Kodak RP 30 manual Prerequisites for MBL and PBL Calibration, Printing a PBL Test, Test Start PBL