Operation Procedure (With data management software CS-S10w)

Setting Necessary for Object Color Measurement

1. Set one or more tungsten lumps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Illumination

 

 

 

 

 

 

 

 

 

 

 

 

or equivalent as illumination

White calibration plate

Light Source

 

 

 

 

 

 

 

 

 

 

 

 

source toward white calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

plate as in the right illustration.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set this instrument vertical

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

to white calibration plate.

 

 

 

45°

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

45°

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Keep the angle between

 

illumination light source and

 

 

white calibration plate at 45˚.

 

2.

 

 

Illumination

 

 

 

Light Source

 

 

 

 

Start up this instrument, PC, and software CS-S10w.

 

White Calibration

3. Set color measuring mode for object color with CS-S10w.

4. Conduct white calibration with CS-S10w.

Measurement of Object

5. Set object on the same position and at the

same angle as white calibration plate.

6. Now, measurement can be done with CS-S10w.

See instruction manual of data management software CS-S10w for details.

Description