Setting Observer

Color matching function for chromaticity calculation is selectable between 2˚ and 10˚.

Setting at the shipment from factory : 2˚ OBS

Operation Procedure

OBSERVER Setting

2

3

1, 7

6

4, 5

1. Press ESC key when menu or target value setting menu appears.

Measurement screen appears on LCD screen.

2. Make sure that [ ] appears on left bottom indicating this instrument has not been set for SHIFT mode.

If [SFT] appears instead, it means that

CH00:<DEFAULT>

 

<ANGL> 1° VIEW

 

Lv

20.80

 

 

 

<SPD> AUTO

cd/m

2

<SYNC> 200.00Hz

x

 

 

 

<LENS>STANDARD

0.4476

 

 

<MEM>AUTO NUM

y

0.4477

 

 

<

MEMORY DATA>

 

 

OOO

:(NO DATA)

 

 

 

 

 

 

M

 

 

 

 

 

 

 

Lv ------

<MODE>SINGLE

 

 

x ------

 

 

y ------

[

]

[

2° ]

 

 

 

 

 

 

instrument is to operate in SHIFT mode. Press SHIFT key to cancel SHIFT mode.

3.

4.

Press MENU key four times.

Menu 4/4 screen appears on LCD screen.

Press ENTER key.

Inversion cursor moves from [OBSERVER] to right to change observer angle.

<MENU>4/4

OBSERVER

2˚ OBS

BREAK : [ESC]

[ ] [

2° ]

<MENU> 4/4

OBSERVER

2˚ OBS

 

 

BREAK : [ESC]

[

]

[ 2° ]

32