Error

Symptom

Item to Check

Corrective Action

Page to

 

 

No.

Refer

 

 

 

 

 

 

 

9

Display of

Is object to measure

When an object whose luminance

 

 

 

 

the remaining

stable?

has changed greatly from the pre-

 

 

 

 

measurement

 

vious measurement is measured,

 

 

 

 

time freezes,

 

measurement may be interrupted to

 

 

 

 

and measure-

 

set the optimum gain for measure-

 

 

 

 

ment is not

 

 

 

 

 

 

ment and then remeasurement will

 

 

 

 

completed

 

 

 

 

 

 

be performed; during remeasure-

 

 

 

 

for the set

 

 

 

 

 

 

ment, the remaining measurement

 

 

 

 

measurement

 

 

 

 

 

 

time display is frozen. Keep object to

 

 

 

 

time.

 

 

 

 

 

 

measure stable.

 

 

 

 

 

 

 

 

 

 

 

Are you measuring an

When an object whose luminance

 

 

 

 

 

object whose lumi-

has changed greatly from the pre-

 

 

 

 

 

nance has changed

vious measurement is measured,

 

 

 

 

 

greatly from the

measurement may be interrupted to

 

 

 

 

 

previous measure-

 

 

 

 

 

set the optimum gain for measure-

 

 

 

 

 

ment?

 

 

 

 

 

ment and then remeasurement will

 

 

 

 

 

 

 

 

 

 

 

 

be performed; during remeasure-

 

 

 

 

 

 

ment, the remaining measurement

 

 

 

 

 

 

time display is frozen.

 

 

 

10

PC cannot

Has USB cable been

Connect this instrument and PC

98

 

 

 

read output

tightly connected?

securely.

 

 

 

 

data from this

 

 

 

 

 

 

instrument

 

 

 

 

 

 

Hasn't remote mode

Switch to remote mode by sending

99

 

 

 

when con-

 

 

 

been cancelled?

connection command from PC to this

 

 

 

 

nected with

 

 

 

 

 

instrument.

 

 

 

 

USB. Com-

 

 

 

 

 

 

Use standard accessory data manage-

 

 

 

 

mand or data

 

 

 

 

 

cannot be

 

ment software CS-S10w Standard.

 

 

 

 

entered from

Has prepared program

Check referring to sample program. Use

-

 

 

 

PC to this

been correct?

standard accessory data management

 

 

 

 

instrument.

 

software CS-S10w Standard.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

Measurement

Hasn't backup battery

Switch ON to charge backup battery.

4

 

 

 

data or vari-

been insufficient after

Backup battery is to fulfill its func-

 

 

 

 

ous setting

a long time period of

tion after fully charged for approx. 20

 

 

 

 

becomes

nonuse? Has backup

hours.

 

 

 

 

invalid soon.

battery been charged

Shelf life of backup battery is 10

 

 

 

 

 

enough at the time of

years under general use, but re-

 

 

 

 

 

purchased?

placement is needed should memo-

 

 

 

 

 

 

ry disappears soon after full charge.

 

 

 

 

 

 

For replacing backup battery, please

 

 

 

 

 

 

contact the nearest KONICA MI-

 

 

 

 

 

 

NOLTA SENSING authorized service

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

 

 

 

facility.

 

 

 

 

 

 

 

 

12

Same error

Check corrective action

If symptom does not improve, please

108

 

 

 

message ap-

for error message.

contact the nearest KONICA MI-

 

 

 

 

pears repeat-

 

NOLTA SENSING authorized service

 

 

 

 

edly.

 

facility.

 

 

 

115