5. Press TARGET key to go to target color entry screen.6. Enter target color in numerical value.

key: 0 to 9 in ascending order. K, M, decimal point and space available. If kept pressed, value continuously changes.

key: 9 to 0 descending order. K, M, decimal point and space available. If kept pressed, value continuously changes.

Kand M show x103 and x106 respectively.

7.

(See p.71 for details on ‘Numerical value entry ranges’ )

Press

key to move the cursor

to second digit position.

8. Repeat procedures from 6. to 7. as necessary to enter target color of Lv, x and y respectively.

Press ESC key to stop.

9. Press ENTER key.

After“ PLEASE WAIT...” appears, entered value has been set as target color.

When“ PLEASE WAIT...” appears, do not switch OFF. Doing so may break stored data.

10. Press ESC key to return to measurement screen.

<TARGET>

 

 

 

 

<ANGL> 1° VIEW

 

 

 

 

CH01:Display*1

 

 

<SPD> AUTO

Lv

------

 

 

 

<SYNC>NO SYNC

 

2

 

<LENS>STANDARD

 

 

 

 

cd/m

 

 

<MEM>AUTO NUM

 

x

------

 

 

 

 

 

 

 

 

 

y

------

 

 

 

 

SET CAL VALUE.

 

 

 

OK : [ENTER]

 

 

 

[ SFT]

[ 2° ]

 

 

 

<TARGET>

 

 

 

 

<ANGL> 1° VIEW

 

 

 

 

CH01:Display*1

 

 

<SPD> AUTO

Lv

9-----

 

 

<SYNC>NO SYNC

2

 

<LENS>STANDARD

 

 

 

 

cd/m

 

 

<MEM>AUTO NUM

 

x

------

 

 

 

 

 

 

 

 

 

y

------

 

 

 

 

SET CAL VALUE.

 

 

 

OK : [ENTER]

 

 

 

[ SFT]

[ 2° ]

 

 

 

<TARGET>

 

 

 

 

<ANGL> 1° VIEW

 

 

 

 

CH01:Display*1

 

 

<SPD> AUTO

Lv

9-----

 

 

<SYNC>NO SYNC

2

 

<LENS>STANDARD

 

 

 

 

cd/m

 

 

<MEM>AUTO NUM

 

x

------

 

 

 

 

 

 

 

 

 

y

------

 

 

 

 

SET CAL VALUE.

 

 

 

OK : [ENTER]

 

 

 

[ SFT]

[ 2° ]

 

 

 

<TARGET>

 

 

 

 

<ANGL> 1° VIEW

 

 

 

 

CH01:Display*1

 

 

<SPD> AUTO

Lv

90.78

 

 

 

 

<SYNC>NO SYNC

 

cd/m

2

<LENS>STANDARD

 

x

------

 

 

<MEM>AUTO NUM

 

 

 

 

 

 

y

------

 

 

 

 

SET CAL VALUE.

 

 

 

OK : [ENTER]

 

 

 

[ SFT]

[ 2° ]

 

 

 

<TARGET>

 

 

 

 

<ANGL> 1° VIEW

 

 

 

 

CH01:Display*1

 

 

<SPD> AUTO

Lv

90.78

 

 

 

 

<SYNC>NO SYNC

cd/m

2

<LENS>STANDARD

 

 

 

 

 

 

<MEM>AUTO NUM

x

0.3545

 

 

 

 

 

 

y

0.3965

 

 

 

SET CAL VALUE.

 

 

 

OK : [ENTER]

 

 

 

[ SFT]

[ 2° ]

 

 

 

CH

 

:<DEFAULT>

 

 

<ANGL> 1° VIEW

01

 

 

Lv

------

 

 

<SPD> AUTO

 

2

<SYNC>NO SYNC

 

x

 

 

cd/m

 

<LENS>STANDARD

 

------

 

 

<MEM>AUTO NUM

 

y

------

 

 

<MEMORY DATA>

 

 

 

 

 

 

 

 

M000 :(NO DATA)

 

 

 

 

 

 

 

 

Lv ------

<MODE>SINGLE

 

 

x ------

 

 

y ------

[ SFT ]

 

[ 2° ]

 

 

 

 

 

TARGET Measurement Preparation

85