Error Messages

Error message appears on LCD screen for incorrect operation of this instrument through key. Below table shows type of error message, its description and corrective action respectively.

 

 

 

Error message

Cause (Description)

Corrective Action

 

 

1

BATTERY OUT

Battery voltage decreases.

• Switch this instrument OFF (O

 

 

 

 

 

side), and replace with new batter-

 

 

 

 

 

ies or use AC adapter.

 

 

 

 

 

 

2

DATA SET ERROR

Entered numerical value is

• Enter again. Note that measure-

 

 

 

 

out of range.

ment value should completely

 

 

 

 

 

satisfy the ranges below:

 

 

 

 

 

0<x<1,

 

 

 

 

 

0<y<1,

 

 

 

 

 

0<x+y1,

 

 

 

 

 

0<X99999000000,

 

 

 

 

 

0<Y99999000000,

 

 

 

 

 

0Z99999000000

 

3

CH00 DATA

When CH00 is selected,

• Select channel except CH00 and

 

 

 

NOT SETTA-

calibration channel is go-

redo the operation.

 

 

 

BLE BY USER

ing to be operated or target

 

 

 

 

 

color is going to be set.

 

 

4

INCORRECT

Measurement data measured

• Select measurement data meas-

 

 

 

OBSERVER

with different observer angle

ured with channel and observer

 

 

 

CONDITION

from that for channel is going

angle or reset calibration channel,

 

 

 

 

to be used for user calibration

and redo setting.

 

 

 

 

for that channel, or for target

 

 

 

 

 

color setting.

 

 

5

NO DATA

Without measurement data,

• Select data to measure or meas-

 

 

 

 

user calibration is going

urement data, and redo setting.

 

 

 

 

to be performed, or target

 

 

 

 

 

color to be set.

 

 

6

UNDER

Luminance of measuring

• Remove lens cap and check if

 

 

 

 

object is lower than avail-

luminance of measuring object

 

 

 

 

able measurement range.

is within available measurement

 

 

 

 

 

range, and then redo measure-

 

 

 

 

 

ment.

 

 

 

 

 

• Check if user calibration is cor-

 

 

 

 

 

rectly performed, and redo meas-

 

 

 

 

 

urement.

 

 

 

 

 

• Check if settings for measurement

 

 

 

 

 

angle, observer angle and lens are

 

 

 

 

 

correct, and redo measurement.

 

 

 

 

 

Description

 

 

 

• If symptom does not improve,

 

 

 

 

 

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 