3. Press MEMORY key.

If [MAN NUM] has been set for memory channel update, measurement value

is to store in selected channel.

In case of [AUTO NUM], measurement value is to store in selected channel and next channel number appears.

Maximum value/minimum value is stored in max/min value measurement. (Note that measurement value for every measurement is stored when [AUTONUM] is set in max/min value measurement.)

If set for [AUTO NUM] or [AUTOSAVE] and all memory channels have been used up, M000 is to use again to be overwritten.

In case that stored data already exists in the channel to store in, warning message "OK TO OVERWRITE?" appears.

If OK, press ENTER key, and if not, ESC key. This warning message can be set

not to appear. (See p.48.)

 

CH01:<DEFAULT>

 

<ANGL> 1° VIEW

 

 

 

 

 

 

<SPD> AUTO

 

 

OK TO OVERWRITE?

 

 

OK

 

: [ENTER]

 

 

 

 

 

CANCEL

 

: [ESC]

 

 

 

 

 

 

<MODE>SINGLE

 

y ------

 

 

[ ] [ 2° ]

 

 

 

 

 

 

CH01:<DEFAULT>

 

<ANGL> 1° VIEW

 

 

 

Lv 20.80

 

 

<SPD> AUTO

 

cd/m

2

<SYNC> 60.00Hz

 

x

 

 

 

<LENS>STANDARD

 

0.4476

 

<MEM>AUTO NUM

 

y

0.4477

 

<

MEMORY DATA>

 

 

001

 

 

 

 

 

 

 

M

:

 

 

 

 

 

 

 

 

Lv

20.80

 

<MODE>SINGLE

 

 

x

0.4476

 

 

 

y

0.4477

 

[

]

[ 2° ]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CH01:<DEFAULT>

 

<ANGL> 1° VIEW

 

 

 

Lv 20.80

 

 

<SPD> AUTO

 

cd/m

2

<SYNC> 60.00Hz

 

x

 

 

 

<LENS>STANDARD

 

0.4476

 

<MEM>AUTO NUM

 

y

0.4477

 

<

MEMORY DATA>

 

 

002

: (NO DATA)

 

 

 

 

 

 

M

 

 

 

 

 

 

 

Lv

------

 

<MODE>SINGLE

 

 

x

------

 

 

 

y

------

 

[

]

[ 2° ]

 

 

 

 

 

 

 

 

 

 

MEMORY Measurement

91