Description

Specification

Item

 

CHROMA METER CS-200

 

 

 

 

 

 

 

Measurement range

0.01 - 200,000 cd/m2

(Measuring angle 1˚)

 

 

 

0.01 - 5,000,000 cd/m2

(Measuring angle 0.2˚)

 

 

0.01 - 20,000,000 cd/m2

(Measuring angle 0.1˚)

 

Accuracy

150 cd/m2

(for illuminant A)

 

L ±2 %±1digit

xy ±0.002

 

 

 

 

 

v

 

 

 

(Measuring angle 1˚)1

0.01 - 0.5 cd/m2

(for illuminant A)

 

L

±0.02 cd/m2±1digit ---

 

 

 

 

 

v

 

 

 

(Temperature: 23˚±2˚,

0.5 - 1 cd/m2

(for illuminant A)

 

L

±0.02 cd/m2±1digit xy ±0.007

 

 

 

 

 

v

 

 

 

RH 65% max.)

1 - 10 cd/m2

(for illuminant A)

 

L ±2 %±1digit

xy ±0.004

 

 

 

 

 

v

 

 

 

 

10 - 200,000 cd/m2 (for illuminant A)

 

L ±2 %±1digit

xy ±0.003

 

 

 

 

 

v

 

 

 

 

(5,000 cd/m2 for illuminant A) color filter (R,G,B)

xy ±0.006

Repeatability

0.01 - 1 cd/m2

(for illuminant A)

L

0.01 cd/m2+1digit ---

(2 σ) (AUTO)

 

 

 

 

v

 

 

 

 

(Measuring angle 1˚)2

1 - 2 cd/m2

(for illuminant A)

L

0.5 %+1digit

xy 0.002

(2 σ) (AUTO)

 

 

 

 

v

 

 

 

 

 

2 - 4 cd/m2

(for illuminant A)

L

0.5 %+1digit

xy 0.001

(2 σ) (AUTO)

 

 

 

 

v

 

 

 

 

 

4 - 8 cd/m2

(for illuminant A)

L

0.5 %+1digit

xy 0.0005 (2 σ) (AUTO)

 

 

 

 

v

 

 

 

 

 

8 - 200,000 cd/m2(for illuminant A)

L

0.1 %+1digit

xy 0.0004 (2 σ) (AUTO)

 

 

 

 

v

 

 

 

 

Measurement time

Automatically set between approx. 1 sec/meas. and 60 sec/meas. (AUTO)

 

Automatically set to approx. 1 sec/meas. or 3 sec/meas. (LTD.AUTO)

 

approx. 0.5 sec/meas.(Super-FAST)

approx. 1 sec/meas.(FAST)

 

approx. 3 sec/meas. (SLOW)

 

approx. 12 sec/meas. (Super-SLOW)

Measurement method

Spectral method, Grating + linear photo diode array

 

 

 

 

 

 

 

 

 

 

Measuring angle

1˚, 0.2˚, 0.1˚ (Switchable)

 

 

 

 

 

 

 

 

 

 

Minimum measuring area

ø0.5 mm, ø0.1 mm (With close-up lens No.107)

 

 

 

 

 

Minimum measuring

296 mm (Distance from front edge of metal lens barrel)

 

distance

 

 

 

 

 

 

 

 

Color space

Lv x y, Lv u’ v’, Lv Tuv, XYZ,

dominant wavelength

 

 

Measurement synchroniza-

Vertical synchronization frequency:40.00 to 200.00 Hz

 

tion frequency

 

 

 

 

 

 

 

 

Interface

USB1.1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power source

AC adapter and 4 AA-Size Batteries

 

 

 

 

 

 

 

Battery life

3 hours approx. (in Continuous measurement/FAST mode, with AA-Size bat-

 

tery under company testing KONICA MINOLTA SENSING's conditions.)

Size

95 (W) x 127(H) x 334(L) mm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight

1.8 kg (without battery)

 

 

 

 

 

 

 

 

 

Operating temperature

0 to 40˚C, RH 85 % or less (at 35˚C) with no condensation

 

/humidity range

 

 

 

 

 

 

 

 

Storage temperature

0 to 40˚C, RH 85 % or less (at 35˚C) with no condensation

 

/humidity range

 

 

 

 

 

 

 

 

Standard accessory

Lens cap, Holding cap CS-A24, ND eyepiece filter CS-A27, AC adapter AC-

 

A23, Data management software CS-S10w Standard, USB cable (2 m) IF-A17

Optional accessory

Close-up lens No.107, Close-up lens No.122, ND filter (1/10) CS-A6,

 

ND filter (1/100) CS-A7, Calibration instruction (For ND filter),

 

 

Step up ring (40.5 to 55 mm) CS-A26, Angle finder VN, White calibration plate

 

(For 45-0) CS-A20, White calibration plate (For d-0) CS-A21, White calibra-

 

tion plate CS-A22, Soft case CS-A23, Data management software CS-S10w

 

Professional

 

 

 

 

 

 

 

1. 23˚C±2˚C, Lv=0.01 to 10 cd/m2 SLOW average of 30 measurements Lv= 10 cd/m2 and higher SLOW average of 10 measurements

2. At 0.2˚ Measuring angle, the amount of received light is approx. 1/25 of that for 1˚. Therefore, the repeatability becomes the same as that for 1˚ with 25 times lower luminance. 0.1˚ Measuring angle, the amount of received light is approx. 1/100 of that for 1˚.

Therefore, the repeatability becomes the same as that for 1˚ with 100 times lower luminance.

120