Implementing User Calibration

User calibration cannot be conducted in calibration channel CH00.

(CH00 serves as calibration channel for measurement based on KONICA MINOLTA SENSING’s calibration standard.)

When user calibration is performed in the calibration channel for which target color has already been set, previous target color is cancelled.

These are commonly used among each color space of Lvxy, Lvu’v’, LvTuv, XYZ, and dominant wavelength in one channel.

Preparation

Measurement

 

 

SETTING

CH

 

 

Operation Procedure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

3

 

 

1,

10

 

4, 9

1. Press ESC key when menu or target value setting menu appears.

Measurement screen appears on LCD screen.

5, 7

5

4, 6

2. Press SHIFT key to switch
to SHIFT mode.

[SFT] appears on left bottom indicating

this instrument is to operate in SHIFT

mode. Or make sure that [SFT] appears on

left bottom indicating that this instrument

has been operated in SHIFT mode.

CH01:<DEFAULT>

Lv

------

x

 

cd/m2

------

y

------

<MODE>SINGLE

[ SFT]

[ 2° ]

<ANGL> 1° VIEW

<SPD> AUTO

<SYNC>NO SYNC

<LENS>STANDARD

<MEM>AUTO NUM

<MEMORY DATA>

M000 :(NO DATA)

Lv ------

x------

y ------

3. Press either

or

key to

switch calibration channel and select one for user calibration.

CH

 

:<DEFAULT>

 

<ANGL> 1° VIEW

01

 

Lv

------

 

 

<SPD> AUTO

cd/m

2

<SYNC>NO SYNC

 

x

 

 

 

<LENS>STANDARD

 

------

 

<MEM>AUTO NUM

 

y

------

 

<MEMORY DATA>

 

 

 

 

 

 

 

M000 :(NO DATA)

 

 

 

 

 

 

 

Lv ------

<MODE>SINGLE

 

x ------

 

y ------

[ SFT ]

[ 2° ]

 

 

 

64