11. Enter value for calibration. key: 0 to 9 in ascending order. K, M,

decimal point and space available. If kept pressed, value continuously changes. key: 9 to 0 descending order. K, M,

decimal point and space available. If kept pressed, value continuously changes.

Kand M show x103 and x106 respectively.

(See p.71 for details on ‘Numerical value entry ranges’ )12. Press key to move cursor to second digit position.

13. Repeat procedures from 10 to 11 as necessary to enter calibration values for Lv, x, and y respectively.

Press ESC key to stop.

<USER CAL DATA> CH01:<DEFAULT>

Lv 1-----

x

cd/m2

------

y

------

SET CAL VALUE.

OK : [ENTER]

[ SFT] [ 2° ]

<USER CAL DATA> CH01:<DEFAULT>

Lv 1-----

x

cd/m2

------

y

------

SET CAL VALUE.

OK : [ENTER]

[ SFT] [ 2° ]

<USER CAL DATA> CH01:<DEFAULT>

Lv 100.77cd/m2 x 0.3945 y 0.4000

SET CAL VALUE.

OK : [ENTER]

[ SFT] [ 2° ]

<ANGL> 1° VIEW

<SPD> AUTO

<SYNC>NO SYNC

<LENS>STANDARD

<MEM>AUTO NUM

<MEMORY DATA>

M002 :

Lv 100.76

x0.3945 y 0.3998

<ANGL> 1° VIEW

<SPD> AUTO

<SYNC>NO SYNC

<LENS>STANDARD

<MEM>AUTO NUM

<MEMORY DATA>

M002 :

Lv 100.76

x0.3945 y 0.3998

<ANGL> 1° VIEW

<SPD> AUTO

<SYNC>NO SYNC

<LENS>STANDARD

<MEM>AUTO NUM

<MEMORY DATA>

M002 :

Lv 100.76

x0.3945 y 0.3998

CAL USER Measurement Preparation

14. Press ENTER key.

After“ PLEASE WAIT...” appears,

<CH SETTING> screen appears.

When“ PLEASE WAIT...” appears, do not switch OFF. Doing so may break stored data.

15. Press ESC key twice to return

to measurement screen.

<CH SETTING>

CH01:<DEFAULT>

USER CAL

CH ID NAME

CH COPY

CH RESET

BREAK:[ESC]

[ SFT]

[ 2° ]

CH

 

:<DEFAULT>

01

Lv

------

 

x

 

cd/m2

 

------

 

y

------

<MODE>SINGLE

[ SFT]

[ 2° ]

<ANGL> 1° VIEW

<SPD> AUTO

<SYNC>NO SYNC

<LENS>STANDARD

<MEM>AUTO NUM

<MEMORY DATA>

M002 :

Lv 100.76

x0.3945 y 0.3998

69