Error message

Cause (Description)

Corrective Action

 

 

7

OVER

Luminance of measuring

• Use ND filter and redo measure-

 

 

 

 

object is higher than avail-

ment.

 

 

 

 

able measurement range.

• If symptom does not improve,

 

 

 

 

 

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 

 

 

8

OFFSET ERROR

Zero calibration has not

• Reset the power and redo meas-

 

 

 

 

been correctly performed.

urement.

 

 

 

 

 

• If symptom does not improve,

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 

 

 

9

EXCESSIVE LUMI-

Luminance change for

• Place this instrument on stable

 

 

 

NANCE VARIATION

measuring object is large.

surface and redo measurement.

 

 

 

 

 

• If symptom does not improve,

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 

 

 

10

VIEWING-ANGLE

Measurement was per-

• Switch measurement angle selec-

 

 

 

SELECTOR ERROR

formed when measurement

tor and check that measurement

 

 

 

 

angle selector was in wrong

angle appears in screen (------ is

 

 

 

 

position, or measurement

not displayed), and then redo

 

 

 

 

angle selector was changed

measurement. In addition, do not

 

 

 

 

during measurement.

operate measurement angle selec-

 

 

 

 

 

tor during measurement.

 

 

 

 

 

• If symptom does not improve,

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 

 

 

11

ROM ERROR

Data stored in ROM is

• Do not switch OFF (O side) while

 

 

 

 

broken.

storing data or changing setting, or

 

 

 

 

 

when message “PLEASE WAIT...”

 

 

 

 

 

 

 

 

 

Description

 

 

 

appears.

 

 

 

 

 

 

 

 

 

• If symptom does not improve,

 

 

 

 

 

please contact the nearest KONI-

 

 

 

 

 

CA MINOLTA SENSING authorized

 

 

 

 

 

service facility.

 

 

 

 

 

 

 

 

111