5. Press MENU key.

Menu 1/4 screen appears on LCD screen.

6. Press either or key to select <CH SETTING> and then ENTER key.< CH SETTING> screen appears.7. Press either or key to select[CH RESET] and then ENTER key.< RESET > screen appears.8.Press ENTER key.

After“ PLEASE WAIT...” appears and it is reset, <RESET> screen appears. When“ PLEASE WAIT...” appears, do not switch OFF. Doing so may break stored data.

9.As in procedure 3, you can switch calibration channel to continue to reset other calibration channel.

Note that original calibration channel

appears when you return to measurement screen in procedure 10.

10.Press ESC key three times to return to measurement screen.

 

 

<MENU>

1/4

 

 

CH SETTING

 

 

 

MEAS MODE

SINGLE

 

SYNC MODE

 

 

DELETE

 

 

 

 

 

 

 

BREAK : [ESC]

 

[

 

]

[ 2° ]

 

 

 

<CH SETTING>

 

 

 

 

 

CH01:<DEFAULT>

 

 

 

 

USER CAL

 

 

 

 

 

 

CH ID NAME

 

 

 

 

 

 

 

 

CH COPY

 

 

 

 

 

 

CH RESET

 

 

 

 

 

 

BREAK:[ESC]

 

 

 

[

 

]

[ 2° ]

 

 

 

<CH SETTING>

 

 

 

 

 

CH01:<DEFAULT>

 

 

 

 

USER CAL

 

 

 

 

 

 

CH ID NAME

 

 

 

 

CH COPY

 

 

 

 

 

 

CH RESET

 

 

 

 

 

 

BREAK:[ESC]

 

 

 

[

 

]

[ 2° ]

 

 

 

<RESET> CH01:

Lv 100.77

x0.3945 y 0.4000

OK TO RESET

THIS CH ?

OK

: [ENTER]

CANCEL : [ESC]

[ ]

[ 2° ]

CH01:<DEFAULT>

 

<ANGL> 1° VIEW

 

Lv

------

 

 

<SPD> AUTO

cd/m

2

<SYNC> 60.00Hz

x

 

 

 

<LENS>STANDARD

------

 

<MEM>AUTO NUM

y

------

 

<

MEMORY DATA>

 

000

:(NO DATA)

 

 

 

 

 

M

 

 

 

 

 

 

Lv ------

<MODE>SINGLE

 

 

x ------

 

 

y ------

[ ]

[ 2° ]

 

 

 

 

 

 

RESET CH Measurement Preparation

73