Sensor Parameters
This facility is password protected. It stores sensor information used by Micronics and is not available for the user.
WARNING! Sensor should only be edited following instruction from the factory Enter password
Factory Settings
The facility is used by Micronics in the process of instrument calibration. Press ENTER takes you back to SETUP INSTRUMENT MENU.
Exit
Means EXIT and will take you back to the Main Menu.
Main Menu - Read Flow
When choosing the Read flow option from the MAIN MENU the instrument reverts directly back to the data that was last entered. The instrument will have to be reprogrammed if it is to be used on a new application.
KEYPAD OPTIONS
The output options can only be adjusted/operated in flow mode.
Exit
Delete Key
If anything is entered in error, press the DELETE key and
Options Key
This can only be used in flow mode. Scroll down the options then press ENTER to select.
OPTIONS |
|
Cutoff (m/s) | 0.05 |
Set zero |
|
Total | RUN |
Reset + total |
|
Reset – total |
|
Damping (sec) | 5 |
Cal Factor | 1000 |
Corr Factor | 1000 |
Diagnostics |
|
Exit |
|
Cut Off (m/s)
The instrument has an automatic CUTOFF that is calculated to 0.05 m/s. The maximum flow is calculated when the instrument is programmed and is displayed when sensor set and mode of operation are displayed (See page 9 - Read Flow - Attach sensors). Micronics cannot guarantee measuring flows below this range because of instabilities in measuring system, but it is possible for the user to cancel any
Reducing the cutoff to Zero allows the user to see or record any flow that they may not want. For example it may be that the user may not want to measure flows below 50 LPM in a 50mm pipe that is equivalent to 0.42 m/sec, in which case 0.42 m/sec would be entered into the instrument and nothing would be recorded below that level. The maximum cut off is 1 m/sec.
13