Olympus M1200 manual Priming the High-PressurePump, Prime Valve

Models: M1200

1 45
Download 45 pages 26.89 Kb
Page 12
Image 12
3. Priming the High-Pressure Pump:

3. Priming the High-Pressure Pump:

Once water is in the solution tank, the high-pressure solution pump must be primed:

There is a prime valve located on the front lower portion of the machine.

Turn on both vacuums and the water pump. Press the priming lever for a few seconds while you block off the vacuum inlet with your hand. The Vacuum will pull solution through the pump and prime valve into the vacuum tank.

Let off the prime valve and your pump is then primed. As long as there is solution in the tank the pump should remain primed.

After priming, turn off the pump and step on the primer valve to relieve the pressure in the line.

If the pump still does not prime, or if flow is low or unsteady, check the hose from the solution tank to the pump (as well as the inline filter) for clogging, kinks, or restrictions. Clean or replace hose and/or filter and repeat the priming procedure.

If you are having trouble with the pump, refer to the trouble shooting guide or contact your distributor for advice or assistance.

Prime Valve

12

Page 12
Image 12
Olympus M1200 manual Priming the High-PressurePump, Prime Valve

FAQ

What is the prime valve on the Olympus M1200 Vacuum Cleaner?

The prime valve is located on the front lower portion of the machine and is used to prime the high-pressure solution pump.

How do I relieve pressure in the line after priming the pump?

After priming, turn off the pump and step on the primer valve to relieve the pressure in the line.

What should I do if the pump still does not prime or if the flow is low or unsteady?

Check the hose from the solution tank to the pump as well as the inline filter for clogging, kinks, or restrictions. Clean or replace hose and/or filter and repeat the priming procedure.