12.Reconnect vacuum hose to vacuum manifold.

10.Apply thread sealant to the threads of the stem retainer and screw the retainer back into the regulator body. Tighten using wrenches.

Apply thread sealant to stem retainer

Screw stem retainer back into regulator body

11.Place the regulator back into the opening on the base wall. Reconnect any hoses which were removed to access the regulator. (Use thread sealant when reconnecting hoses.)

12.Reconnect vacuum hose to vacuum manifold.

Vacuum Hose – Connected to Vacuum Manifold

Regulator – Inserted into opening on base wall

13.Close tanks back onto base and secure latch.

14.Place washer and jam nut back onto shoulder of regulator and tighten to secure regulator to base wall.

15.Replace spring washer, spring, and spring plate back onto regulator stem.

16.Screw black knob back onto regulator stem.

17.Screw nut back onto regulator stem. Do not screw nut on too far, as the nut may interfere with the movement of the regulator knob and prevent the knob from being backed out when trying to adjust regulator to lower pressures.

32

Page 32
Image 32
Olympus M1200 manual Reconnect vacuum hose to vacuum manifold

FAQ

What thread sealant should be used for the stem retainer?

It is recommended to use thread sealant on the threads of the stem retainer to ensure a secure fit and prevent leaks.

How should I reconnect the vacuum hose to the vacuum manifold?

Simply reconnect the vacuum hose to the vacuum manifold to restore proper suction and functionality.

What precautions should be taken when screwing the nut back onto the regulator stem?

Do not screw the nut on too far to prevent interference with the movement of the regulator knob.

Can I use any wrench to tighten the stem retainer back into the regulator body?

It is recommended to use appropriate wrenches to ensure proper tightening without damaging the components.

How to reattach the regulator to the base wall?

Simply place the regulator back into the opening on the base wall and reconnect any hoses that were removed.

What is the purpose of the washer and jam nut on the shoulder of the regulator?

The washer and jam nut are used to secure the regulator to the base wall.

Is it necessary to use thread sealant when reconnecting hoses?

Yes, it is advisable to use thread sealant when reconnecting hoses to prevent any potential leaks.

How to adjust the regulator to lower pressures?

To adjust the regulator to lower pressures, you may need to back out the regulator knob while ensuring that the nut is not screwed on too far.

What components should be replaced onto the regulator stem?

Replace the spring washer, spring, and spring plate back onto the regulator stem to ensure proper functionality.

What is the importance of securing the latch when closing tanks back onto the base wall?

Securing the latch ensures that the tanks are properly closed and sealed on the base wall to avoid any potential leaks.

What are the steps to reconnect the vacuum hose to the vacuum manifold?

The steps to reconnect the vacuum hose to the vacuum manifold are clearly outlined in this article with detailed instructions.