Polycom DOC2559A manual Rtm Isdn/Rtm

Models: DOC2559A

1 65
Download 65 pages 53.98 Kb
Page 39
Image 39

Polycom RMX 4000 Hardware Guide

Table 1-7RMX™ 4000 Slot Numbering

Slot ID/No.

Card/Component

Requirement

 

 

 

7

Logo Panel

Not Available (NA)

 

 

 

8

CTNL 4000 unit

Mandatory

 

(CPU 1)

 

 

 

 

9-11

AC Power Supply

An RMX with AC power has 3 power

 

 

supplies installed. A 3rd power supply is

 

 

redundant (n+1).

 

 

Note: Not used with DC powered

 

 

systems. DC powered systems receive

 

 

Direct Current from the power rail .

 

 

 

12

Fan Drawer

Mandatory

 

 

 

13-16

RTM ISDN/RTM

Either an RTM ISDN or an RTM LAN

 

LAN

card is mandatory in combination with a

 

 

Media card. The RTM ISDN/RTM LAN

 

 

board must be inserted in a slot opposite

 

 

any MPM+/MPMx card.

 

 

 

17

RTM-IP 4000

Mandatory

 

 

 

18

Blank Panel

Not Available (NA)

 

 

 

19-21

Power Modules

Mandatory:

 

 

With AC power, 3 power supplies are

 

 

installed, with the 3rd redundant.

 

 

With DC power, 2 power supplies are

 

 

installed with the 2nd redundant. The

 

 

center slot (#20) on the rear of the RMX

 

 

4000 is disabled and is fitted with blank

 

 

panel.

 

 

Note: Protective bonding conductor size

 

 

is 14AWG (1.5mm) within the Power

 

 

Entry Model.

 

 

 

1-33