START-UP OF UNIT

START-UP OF UNIT

Follow the procedures for the initial and consequent operations of the unit.



1.Connect the power cord to the dedicated outlet with appropriate rating.

Manual backgroundCAUTION

If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor.

2.On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the alarm buzzer stop key (BUZZER).

3.Set the chamber temperature to 5oC.

4.Allow the chamber temperature to fall to 5oC. Check the chamber temperature on the temperature indicator.

5.Turn on the fluorescent light switch to check the light. After checking, turn off the switch if the light is not necessary.

6.Set the desired temperature. When the chamber temperature gets to the set temperature, begin slowly placing items into the chamber to minimize the temperature rise.

Manual backgroundCAUTION

The chamber is refrigerated by the forced circulation of cooled air inside the chamber. Ensure that the intake and exhaust air vents are not blocked. Adequate space should be provided between the items inside the unit to allow air circulation.

<Attention>

The outer door has a condensation on the glass during refrigerator operation. Wipe off the condensation with a dry soft cloth. Also, wipe off the water on the dew receiver regularly.