![ALARM TEMPERATURE SETTING](/images/new-backgrounds/62071/6207135x1.webp)
ALARM TEMPERATURE SETTING
This unit is provided with both high and low temperature alarms. The temperature at which the alarm is activated may be changed.
The available set range for high temperature alarm is between +2oC and +14oC and
Note: The temperature alarm is set at 5oC of the set temperature at the factory.
Display | Mode | Function |
F01 | High temperature alarm set | See Table 2 on page 17 |
F02 | Low temperature alarm set | See Table 3 on page 17 |
As an example, Table 2 shows the procedure to set the high temperature alarm so that the alarm can activate when the chamber temperature is 3oC higher than the set temperature.
Table 3 shows the procedure to set the low temperature alarm so that the alarm can activate when the chamber temperature is 3oC lower than the set temperature.
Table 2 Procedure for setting high temperature alarm
| Description of operation | Key operated | Indication after operation | ||||||||||||||||||
1 |
| The current chamber temperature is |
|
|
|
|
|
|
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
| |||||||||||
|
| displayed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |||||||||||
2 | Press numerical value shift key for |
| The first digit is flashed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
5 seconds. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
3 | Press numerical value shift key and |
| When pressed, the figure of settable |
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
scroll the figure to 6. |
| digit increases. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
4 | Press set key. | SET | The first digit is flashed. |
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Set the temperature to 03 with the |
| Pressing the key shifts the digit which | ||||||||||||||||||
|
| can be set. | |||||||||||||||||||
5 | digit shift key and numerical value |
| |||||||||||||||||||
| When pressed, the figure of settable |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| shift key. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| digit increases. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
6 | Press set key. | SET | Alarm temperature is memorized and |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
the current chamber temperature is |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
| displayed. | ||||||||||||||||||
Table 3 Procedure for setting low temperature alarm |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| Description of operation | Key operated | Indication after operation | ||||||||||||||||||
1 |
| The current chamber temperature is |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
| displayed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
2 | Press numerical value shift key for |
| The first digit is flashed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
5 seconds. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3 | Press numerical value shift key and |
| When pressed, the figure of settable |
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
scroll the figure to 7. |
| digit increases. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
| |||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
4 | Press set key. | SET | The first digit is flashed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Set the temperature to |
| Pressing the key shifts the digit which | ||||||||||||||||||
|
| can be set. | |||||||||||||||||||
5 | digit shift key and numerical value |
| |||||||||||||||||||
| When pressed, the figure of settable |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
| shift key. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
| digit increases. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
| |||||||||||||||
6 | Press set key. | SET | Alarm temperature is memorized and |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
the current chamber temperature is |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
|
|
| displayed. |
17
- 55 -