OPTIONAL COMPONENTS
Setting for battery before starting
Always perform the setting shown below after the installation of battery for power failure alarm, an optional component
The chamber temperature and “E09” is displayed on the temperature display alternately if the battery switch is turned on without following procedure.
㩵Turn off the battery switch, perform the following setting, and turn on the battery switch again if “E09” is displayed on the temperature display.
| Description of operation | Key operated | Indication after operation | |||||||
1 |
| The current chamber temperature is |
|
| ||||||
|
|
|
|
|
|
| ||||
| displayed. |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
| ||
2 | Press numerical value shift key for |
| F00 is displayed and the first digit |
|
|
|
| |||
|
|
| ||||||||
5 seconds. |
| blinks. |
|
|
|
|
| |||
|
|
|
|
|
|
| ||||
3 | Press numerical value shift key |
| The display is changed to F02 from |
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |||
twice. |
| F00. |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
| |
4 | Press digit shift key once. |
| The second digit blinks. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
5 | Press numerical value shift key 4 |
| The display is changed to F42 from |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
times. |
| F02. |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| ||
6 | Press set key. | SET | The current setting (000) is displayed |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
and the first digit blinks. |
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
| |
7 | Press numerical value shift key and |
| The display is changed to 001 from |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
scroll the figure to 1. |
| 000. |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| ||
|
|
| The setting is memorized and |
|
|
|
|
|
|
|
8 | Press set key. | SET | the current chamber temperature is |
|
|
|
|
|
|
|
|
|
| displayed. |
37
- 75 -