CHAMBER TEMPERATURE SETTING

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is 10oC.

Note: The unit is set at the factory with a chamber temperature of 5oC.

Table 1 Basic operation sequence (Example: Chamber set temperature 10oC)

 

Description of operation

Key operated

Indication after operation

1

Connect the power supply plug.

----

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Press set key.

SET

The second digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set to 010 with the numerical value

 

By pressing key, the settable digits

 

 

 

 

 

 

 

 

 

3

 

moves.

 

 

 

 

 

 

 

shift key and digit shift key.

 

When pressed, the figure of settable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

digit increases.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set temperature is memorized and the

 

 

 

 

 

 

 

 

 

 

 

 

4

Press set key.

SET

current chamber temperature is

 

 

 

 

 

 

 

displayed.

Note:

The temperature set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, the chamber temperature setting is not accepted.

The available set range of chamber temperature is between 2 and 23oC. The partial freezing may occurred if the set temperature is lower than 3oC.