OPTIONAL COMPONENTS
3.Remove the black rubber caps (inside and outside) on the access port located at the top corner of the unit and take the insulation out of the access port.
4.Pass the recorder sensor into the chamber through the access port. (Fig. 3)
5.An attached binder (two pcs) is used for the fan cover of the top of the chamber as the Fig. 4, and a temperature sensor is fixed.
4.Route the capillary on the top of the chamber and fix it with the provided clip and screw. (Fig. 5)
5.Make a cut on the rubber caps so that the capillary can pass the caps and place them back on the access port (inside and outside). (Fig. 6)
9.Bind the lead wire on the right upper side of the unit with the provided band.
10.Remove the connector cap located on the right side and then connect the recorder power. (Fig. 7)
11.Attach the recorder on the front cover with a screw. (Fig. 8)
Temperature recorder sensor
Fig. 3
Binder
Fig. 4
Clip |
Capillary |
Fig. 5 |
Capillary |
Rubber cap
Connector for recorder power
Space for temperature recorder | Fig. 7 |
Fig. 6
Screw |
Shaft |
Fig. 8 |
Channel |